Z-0355-2012

SoftPath GUI Versions: 4.3.0.10, 4.3.0.11, and 4.3.0.12. SoftPath is a laboratory information system to be...

SCC Soft Computer / initiated 2011-06-13 / Terminated / root cause: design

SCC Soft Computer began a recall of SoftPath GUI Versions: 4.3.0.10, 4.3.0.11, and 4.3.0.12. SoftPath is a laboratory information system to be used in a clinical laboratory by knowledgeable... on . The reason FDA records is "On 06/13/2011, SCC Soft Computer initiated a correction on the following SoftPath GUI Versions: 4.3.0.10, 4.3.0.11, and 4.3.0.12. A client reported when they verified the contents of the fax reports, they found one of...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0355-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
SCC Soft Computer
Product
SoftPath GUI Versions: 4.3.0.10, 4.3.0.11, and 4.3.0.12. SoftPath is a laboratory information system to be used in a clinical laboratory by knowledgeable, trained, and experienced personnel. SoftPath enables the user to document steps in the receiving and processing of surgical, gynecological, medical cytological, and autopsy case specimens within the Pathology labs, for the purpose of generating Pathology reports for requesting physicians and label creation for identification of specimen/case material.
Product code
LNX
Reason for recall
On 06/13/2011, SCC Soft Computer initiated a correction on the following SoftPath GUI Versions: 4.3.0.10, 4.3.0.11, and 4.3.0.12. A client reported when they verified the contents of the fax reports, they found one of the specimen diagnoses to be missing. Parts A through D were present as well as Part F, however, Part E diagnosis was missing for one of the patient reports. All affected clients wer
Root cause tag
design
FDA root cause
Device Design
Action
SCC Soft Computer initiated a Field Correction on June 13, 2011. All customers with affected software have been notified of this issue. SCC recommends until the corrected patch is loaded to temporarily use one of the following alternative working solutions: 1. Use MS Word Editor functionality within SoftPath application for creating patient reports. In order to ensure Text Control editor is not inadvertently activated, the templates for Text Control editor will be deleted and reinstalled at the clients request by their SCC representative after receipt of the correction. 2. In Report Template set-up, the bottom margin should be made slightly larger than the size of the footer plus its distance from the bottom page border. For further explanation, please consult your SCC representative for further details, or call (727) 789-0100.
Quantity in commerce
14
Distribution
Class II Recall - Worldwide Distribution -- USA (nationwide) and the country of Canada.
Product codes and lots
SoftPath GUI Versions: 4.3.0.10, 4.3.0.11, and 4.3.0.12.