SoftPath GUI versions: 4.2.3.3-4.2.3.6 and 4.3.0.5-4.3.0.12. SoftPath is a laboratory information system to...
SCC Soft Computer began a recall of SoftPath GUI versions: 4.2.3.3-4.2.3.6 and 4.3.0.5-4.3.0.12. SoftPath is a laboratory information system to be used in a clinical laboratory by knowledgeable... on . The reason FDA records is "On 07/20/2011 SCC Soft Computer, Clearwater, FL, initiated a correction on the following SoftPath GUI versions: 4.2.3.3-4.2.3.6 and 4.3.0.5-4.3.0.12. A client reported that the HPV results on a patient report were...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0363-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- SCC Soft Computer
- Product
- SoftPath GUI versions: 4.2.3.3-4.2.3.6 and 4.3.0.5-4.3.0.12. SoftPath is a laboratory information system to be used in a clinical laboratory by knowledgeable, trained, and experienced personnel. SoftPath enables the user to document steps in the receiving and processing of surgical, gynecological, medical cytological, and autopsy case specimens within the Pathology labs, for the purpose of generating Pathology reports for requesting physicians and label creation for identification of specimen/case material.
- Product code
- JQP Calculator/Data Processing Module, For Clinical Use
- Reason for recall
- On 07/20/2011 SCC Soft Computer, Clearwater, FL, initiated a correction on the following SoftPath GUI versions: 4.2.3.3-4.2.3.6 and 4.3.0.5-4.3.0.12. A client reported that the HPV results on a patient report were incorrect. The HPV results were from the previous order on the patient and not the current order. All affected clients were notified about the issue on 07/20/2011.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- SCC Soft Computer sent a "CORRECTION COMMUNICATION" dated July 20, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. SCC personnel will review the customized Lab Results AP macro and/or the Micro Results AP macro on customer systems and will either discuss with them the irregularities requiring change or reconfigure macros as necessary.
- Quantity in commerce
- 97
- Distribution
- Worldwide Distribution-USA (nationwide) and the country of Canada.
- Product codes and lots
- SoftPath GUI versions: 4.2.3.3-4.2.3.6 and 4.3.0.5-4.3.0.12.