SoftMic GUI versions: 4.0.3.0-4.0.3.15, 4.0.4.1-4.0.4.7. SoftMic is a laboratory information system to be...
SCC Soft Computer began a recall of SoftMic GUI versions: 4.0.3.0-4.0.3.15, 4.0.4.1-4.0.4.7. SoftMic is a laboratory information system to be used in medical research, clinical, and public health... on . The reason FDA records is "On 08/25/2011, SCC Soft Computer initiated a correction on the following SoftMic GUI versions: 4.0.3.0-4.0.3.15, 4.0.4.1-4.0.4.7. For clients using SoftMic Result Entry, when a certain series of keystrokes on the test...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0466-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- SCC Soft Computer
- Product
- SoftMic GUI versions: 4.0.3.0-4.0.3.15, 4.0.4.1-4.0.4.7. SoftMic is a laboratory information system to be used in medical research, clinical, and public health laboratories by knowledgeable, trained, and experienced personnel. SoftMic enables the user to handle data processing and laboratory activities for microbiology and serology testing, for the purpose of storing, managing, querying, and reporting laboratory data.
- Product code
- JQP Calculator/Data Processing Module, For Clinical Use
- Reason for recall
- On 08/25/2011, SCC Soft Computer initiated a correction on the following SoftMic GUI versions: 4.0.3.0-4.0.3.15, 4.0.4.1-4.0.4.7. For clients using SoftMic Result Entry, when a certain series of keystrokes on the test grid are performed while entering results and statusing a test, the test status may not be maintained upon saving. The status may be removed or copied from another test on the order
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- SCC Soft Computer initiated a Field Correction dated August 25, 2011. to all affected customers. Customers with the affected software have been notified and are scheduling receipt of the hot fix.. Further questions call (727) 789-0100.
- Quantity in commerce
- 144
- Distribution
- Worldwide Distribution -- USA (nationwide) and Canada.
- Product codes and lots
- SoftMic GUI versions: 4.0.3.0-4.0.3.15, 4.0.4.1-4.0.4.7.