Maxlock Extreme Extremity Plating System with Variable Angle Technology
Orthohelix Surgical Designs Inc began a recall of MAXLOCK EXTREME EXTREMITY PLATING SYSTEM WITH VARIABLE ANGLE TECHNOLOGY. Product Usage: Variable Angle technology in MTP, TMT, MFT, MXL, DFX, and EDL plates... on . The reason FDA records is "The firm discovered a testing protocol error, in both dynamic and static cantilever comparative testing, during the initial comparative testing". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0421-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- TORNIER Inc.
- Product
- MAXLOCK EXTREME EXTREMITY PLATING SYSTEM WITH VARIABLE ANGLE TECHNOLOGY. Product Usage: Variable Angle technology in MTP, TMT, MFT, MXL, DFX, and EDL plates are indicated for fractures, fusions and osteotomies for small bones in the hand, wrist, foot and ankle in both pediatric and adult patients. Variable Angle Technology in MXL and CLA plates are indicated for the fractures, fusions, and osteotomies of the clavicle and small bones in the hand, wrist, foot and ankle.
- Product code
- HRS Plate, Fixation, Bone
- Reason for recall
- The firm discovered a testing protocol error, in both dynamic and static cantilever comparative testing, during the initial comparative testing.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Orthohelix Surgical Designs Inc. sent an URGENT RECALL NOTIFICATION letter dated August 22, 2011 to all affected customers. The letter identified the affected products, the problem, and the actions to be taken. Customers were instructed to immediately inventory, remove and return the recalled product to the firm via FedEx or any shipping method with tracking capability. Customers were also requested to notify any affected user facilities of the recall. For any questions call 330-247-1444 or 330-247-1445.
- Quantity in commerce
- 2,722 units
- Distribution
- Nationwide Distribution (USA) - including the states of: CA, CO, CT, FL, GA, IA, IL, KY, LA, MD, MI, MO, NC, NJ, NM, NY, OH, OK, PA, SD, TN, TX, UT, VA, WA, WI and WY.
- Product codes and lots
- Lot # 1364, 1731, 1862, Model # VAR-031-35-10, Model # VAR-031-35-12, Model # VAR-031-35-14, Model # VAR-031-35-16, Model # VAR-031-35-18, Model # VAR-031-35-20, Model # VAR-031-35-22, Model # VAR-031-35-24, Model #, VAR-031-35-26, Model # VAR-031-35-28, Model # VAR-031-35-30 & Lot # 1365011, 1365012, Model #, MXL-032-35 and Model # VAR-032-35