Z-0842-2012

Terumo Cardiovascular Systems CAPIOX (r) SX18 Hollow Fiber Oxygenator LOT: HF22 . 2010-04 The X-coated Capiox...

Terumo Cardiovascular Systems Corporation / initiated 2010-06-28 / Terminated / root cause: manufacturing process

Terumo Cardiovascular Corporation began a recall of Terumo Cardiovascular Systems CAPIOX (r) SX18 Hollow Fiber Oxygenator LOT: HF22 . 2010-04 The X-coated Capiox SX18 Hollow Fiber Oxygenators with Detachable... on . The reason FDA records is "Complaints were received of expired X-coated Capiox SX18 Hollow Fiber oxygenators". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0842-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Terumo Cardiovascular Systems CAPIOX (r) SX18 Hollow Fiber Oxygenator LOT: HF22 . 2010-04 The X-coated Capiox SX18 Hollow Fiber Oxygenators with Detachable Hardshell Reservoir are used to exchange gases between extracorporeal blood and a gaseous environment during cardiopulmonary bypass procedures. The integral heat exchanger is used to warm or cool the blood or perfusion fluid and the detachable hardshell reservoir is used to store blood during extracorporeal circulation from both the venous line and the cardiotomy line. The reservoir contains a venous section that is comprised of a filter and defoamer to facilitate air bubble removal.
Product code
DTZ Oxygenator, Cardiopulmonary Bypass
Reason for recall
Complaints were received of expired X-coated Capiox SX18 Hollow Fiber oxygenators.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
All customers were contacted by telephone between 6/28/10 and 7/2/10. The manufacturer mailed "Urgent Medical Device Removal " letters to customers on 01/16/12 with receipt confirmation requested. Customers were informed that expired product had been shipped. Use of expired product was not recommended and product return was requested for refund. Contact Terumo CVS Customer Service 1-800-521-2818 for questions.
Quantity in commerce
16 units
Distribution
Nationwide Distribution including states of: WI, MO,and KY.
Product codes and lots
Product code 3CXSX18RX, Lot # HF22