Z-0511-2012

Urine Reagent Strips (URS) 10 Parameters URS-l0-100W 510(K) number: K970250 Devise Listing number: D072377...

Teco Diagnostics / initiated 2011-10-14 / Terminated / root cause: manufacturing process

Teco Diagnostics began a recall of Urine Reagent Strips (URS) 10 Parameters URS-l0-100W 510(K) number: K970250 Devise Listing number: D072377 Each device consists of 100 strips of Urine Reagent... on . The reason FDA records is "This recall was initiated in response to findings during a FDA inspection of the firm. The firm's recall has been initiated due to improper manufacture of sub-component nitrite paper Lot 35134, Expiration Date: 01/2014...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0511-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Teco Diagnostics
Product
Urine Reagent Strips (URS) 10 Parameters URS-l0-100W 510(K) number: K970250 Devise Listing number: D072377 Each device consists of 100 strips of Urine Reagent Strips 10 parameters packed with a desiccant pouch inside a gray bottle with Teco Diagnostics Label on the bottle and the bottle placed inside a Teco Diagnostics Box. For the semi-quantitative and qualitative detection of Glucose, Bilirubin, Ketone, Specific Gravity, Blood, pH, Protein, Urobilinogen, Nitrite, Leukocytes in urine.
Product code
KQO Automated Urinalysis System
Reason for recall
This recall was initiated in response to findings during a FDA inspection of the firm. The firm's recall has been initiated due to improper manufacture of sub-component nitrite paper Lot 35134, Expiration Date: 01/2014. Discrepancies were found on the information recorded on formulation sheet and manufacturing log sheets (Refer to next paragraph). Sampling size submitted for QC testing was insuff
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Teco Diagnostics sent an "URGENT: MEDICAL DEVICE RECALL" letter dated October 14, 2011 to affected customers. The letter described the product, problem, and actions to be taken by the customers. The letter instructed customers to examine their inventory for the affected product and to quarantine any product found. A Recall Return Response Form was included with the letter for customers to complete and return. Contact the firm's Technical Support Department at (800) 222-9880 for questions regarding this recall.
Quantity in commerce
550 kits
Distribution
Distributed in CA and Poland.
Product codes and lots
Product/Catalog # 1: URS-l 0-100W i. Lot Number: 35264 ii. Expiration Date: 01/2014