CDI Blood Parameter Monitoring System 500
Terumo Cardiovascular Systems Corporation began a recall of CDI Blood Parameter Monitoring System 500 Product Usage: The CDl Blood Parameter Monitoring System 500 is intended for use during cardiopulmonary bypass... on . The reason FDA records is "In a recent review of in-house Service Procedures for the CDI Blood Parameter Monitoring System 500, Terumo Cardiovascular Systems (Terumo CVS) found that one of the procedures it had been performing on older systems...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0673-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- CDI Blood Parameter Monitoring System 500 Product Usage: The CDl Blood Parameter Monitoring System 500 is intended for use during cardiopulmonary bypass procedures when continuous monitoring of arterial and/or venous pH, PC02 , P02, K+, oxygen saturation, hematocrit, hemoglobin, and temperature are desired.
- Product code
- DRY Monitor, Blood-Gas, On-Line, Cardiopulmonary Bypass
- Reason for recall
- In a recent review of in-house Service Procedures for the CDI Blood Parameter Monitoring System 500, Terumo Cardiovascular Systems (Terumo CVS) found that one of the procedures it had been performing on older systems should have been considered a recall at the time it was implemented: Terumo CVS began installing a protective shield (plastic grommet) in the system chassis during manufacture of all
- Root cause tag
- manufacturing process
- FDA root cause
- Component change control
- Action
- Terumo sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated December 6, 2011. The letter identified the affected products, the problem, and the actions to be taken by customer/user. The letter states that Terumo CVS will update all affected CDl Monitors and no additional actions are necessary for those units previously returned that received the correction as part of the service process. For questions or concerns contact Terumo CVS Customer Service 1-800-521-2818.
- Quantity in commerce
- 451 units total
- Distribution
- Worldwide Distribution - USA (nationwide in the states of: AR, AZ, CA, FL, KS, IL, IN, LA, MA, MO, MI, MS, NJ, NV, NC, OH, PA, TN, TX, VA, WA, WI, WV including Puerto Rico and the countries of: AE, Australia, BR, Belgium, Brazil, Canada, China, Dominican Republic, EC, Germany, Guatemala, Japan, Korea, Mexico, SA, Taiwan and UY.
- Product codes and lots
- Catalog Number: 500AVHCT CDI Blood Parameter Monitoring System 500 with Arterial and Venous blood parameter modules and Hematocrit/Saturation probe.