Z-0342-2012
Elekta Synergy XVI X-ray Volume Imaging System To be used as part of radiation therapy treatment process
Elekta Limited / initiated 2011-10-20 / Terminated / root cause: manufacturing process
Elekta, Inc. began a recall of Elekta Synergy XVI X-ray Volume Imaging System To be used as part of radiation therapy treatment process on . The reason FDA records is "A small number of sites have experienced situations with kV generator transformers overheating due to a diode failure on the interface board". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0342-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- Elekta Synergy XVI X-ray Volume Imaging System To be used as part of radiation therapy treatment process.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- A small number of sites have experienced situations with kV generator transformers overheating due to a diode failure on the interface board.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The firm, Elekta, issued a "FIELD CHANGE ORDER MANDATORY ACTION" notice dated October 20, 2011 to its customers to add fuse protection for the kV transformers. The notice describes the product, problem, provides instructions and procedures and actions to be taken. The customers were instructed to follow any instructions or recommendations covered in the notice and to file the document in the Important Notice section of the appropriate User Manual. Elekta will contact the customer if the FCO is applicable to their kV generator. If you have any questions, call 770-300-9725.
- Quantity in commerce
- 123
- Distribution
- Nationwide distribution: USA including states of: AL, AK, AZ, CA, CT, FL, GA, IL, IA, KY, MD, MS, MI, MN, MS, MO, MT, NE, NH, NJ, NY, NC, OH, OK, OR, PA, RI, SC, TX, VT, VA, WA, WV, WI, Puerto Rico, & DC.
- Product codes and lots
- G30369, G29103, G34727, G32313, G30367, G29101, G31330, G34401, G32304, G30962, G30610, G30668, G29986, G28427, G28581, G28673, G28430, G28755, G28576, G28574, G28676, G28678, G28811, G28763, G28806, G28863, G28809, G28864, G28936, G29035, G29042, G30720, G28929, G29043, G29102, G28421, G29296, G29299, G29843, G31188, G29832, G29838, G29844, G29840, G29983, G30300, G30027, G29980, G30210, G30213, G30034, G30208, G30309, G30310, G30530, G30533, G30558, G30565, G32986, G30606, G30611, G30961, G30718, G30662, G30717, G31397, G31326, G31186, G31190, G31323, G29842, G34393, G31404, G31512, G31522, G31518, G31515, G31517, G30073, G32359, G32363, G32305, G32364, G32366, G32385, G32425, G32389, G32501, G32680, G32682, G32809, G32814, G34285, G32905, G32820, G32821, G32824, G32904, G32908, G33335, G33339, G33338, G33428, G33433, G34150, G34288, G34151, G34159, G34290, G34348, G34347, G34350, G34397, G34501, G34684, G34502, G34505, G34689, G30974, G34691, G34980, G34983, G34726