Tobramycin Calibrators, REF 7F93-01; Microgenics Corporation, 46360 Fremont Blvd Fremont, CA 94538...
Microgenics Corp began a recall of Tobramycin Calibrators, REF 7F93-01; Microgenics Corporation, 46360 Fremont Blvd Fremont, CA 94538; Distributed by Abbott Diagnostics, 100 Abbott Park Road... on . The reason FDA records is "Reports that third party control material for 4 lots of Tobramycin Calibrators that are recovering low and outside of the published range". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0773-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Microgenics Corporation
- Product
- Tobramycin Calibrators, REF 7F93-01; Microgenics Corporation, 46360 Fremont Blvd Fremont, CA 94538; Distributed by Abbott Diagnostics, 100 Abbott Park Road, Abbott Park, Illinois 60064. .Tobramycin is a commonly used antibiotic to treat bacterial infections, in particular gram negative infections.
- Product code
- DLJ Calibrators, Drug Specific
- Reason for recall
- Reports that third party control material for 4 lots of Tobramycin Calibrators that are recovering low and outside of the published range.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- ThermoFisher Scientific sent a product removal letter dated November 11, 2011. Letters were sent to the sole distributor of the product. The letter identified the product the problem and the action needed to be taken by the customer. If you are using or have inventory of the affected lot numbers immediately discontinue use and destroy any remaining affected lot numbers per your local waste ordinances. If you have any further questions regarding this information please call 1-877-4Abbott. Customers outside the US should contact their local customer service.
- Quantity in commerce
- 1366 kits
- Distribution
- Worldwide Distribution -- Nationwide Distribution including the country Europe.
- Product codes and lots
- Lots: 59281991, Exp 11-30-11, 59359468, Exp 2-29-12, 59590335, Exp 6-30-12, 59511648, Exp 8-31-12.