CDI Blood Parameter Monitoring System 500, 500AHCT. Intended for use during cardiopulmonary bypass procedures
Terumo Cardiovascular Systems Corporation began a recall of CDI Blood Parameter Monitoring System 500, 500AHCT. Intended for use during cardiopulmonary bypass procedures on . The reason FDA records is "Replacement of the SBC batteries (system batteries) should be managed as "routine maintenance" instead of as an in-house Service Procedure because some systems may be relying on batteries beyond their useful life". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1379-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- CDI Blood Parameter Monitoring System 500, 500AHCT. Intended for use during cardiopulmonary bypass procedures.
- Product code
- DRY Monitor, Blood-Gas, On-Line, Cardiopulmonary Bypass
- Reason for recall
- Replacement of the SBC batteries (system batteries) should be managed as "routine maintenance" instead of as an in-house Service Procedure because some systems may be relying on batteries beyond their useful life.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling design
- Action
- On 2/3/12 all consignees received an URGENT MEDICAL DEVICE RECALL letter via Federal Express. The letter described the reason for the recall, identified the affected product population, what to do in the event of a failure, the potential hazard, and correction. Terumo CVS will update the Operator's Manual for the CDI 500 with instructions to replace the battery within its useful life. Users are to review the notice and ensure that all users are made aware of the issue. In addition, the Customer Response Form should be completed and returned. Users can contact their local Terumo Cardiovascular Systems representative or call 1-800-521-2818 with questions or concerns, to check the last time system batteries were replaced, or to schedule maintenance for systems approaching 10 years.
- Quantity in commerce
- 3,829 units
- Distribution
- Worldwide Distribution -- USA, including the states of AK, AL, AR, AZ, CA, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, WV, WA, and WI and the countries of Japan, Germany, Barbados, Saudi Arabia, Columbia, Uruguay, Ecuador, Trinidad, Costa Rica, Paraguay, Panama, Belgium, Jordan, Libya, Mexico, Brazil, Honduras, Canada, Korea, Dominica Republic, El Salvador, Guatemala, Vietnam, Chile, Hong Kong, United Arab Emirates, Singapore, Australia, Malaysia, Taiwan, Philippines, Mexico, Thailand, Vietnam, and South Africa.
- Product codes and lots
- 500AHCT CDI Blood Parameter Monitoring TBD System 500 with Arterial blood parameter module and Hematocrit/ Saturation probe: 1023, 1025-1027, 1029-1031, 1034, 1035, 1037, 1040, 1045, 1048, 1054-1058, 1060-1062, 1064, 1065, 1072-1074, 1076-1095, 1098-1100, 1102, 1106-1110, 1114, 1116-1129, 1131, 1132, 1141-1145, 1153-1165, 1169-1180, 1186-1191, 1194-1210, 1212-1215, 1221-1230, 1236-1243, 1249-1253, 1256-1260, 1265-1269, 1275-1284, 1290-1294, 1300-1303, 1310, 1311, 1316-1320, 1322-1331, 1335-1339, 1345-1349, 1359-1378, 1384-1388, 1394-1398, 1409-1413, 1419, 1434-1438, 1444-1448, 1454-1458, 1469-1478, 1481, 1482, 1484-1493, 1495-1503, 1505-1520, 1528, 1531-1553, 1555-1558, 1567-1596, 1607-1616, 1632-1636, 1647-1656, 1667-1671, 1677-1681, 1687-1696, 1702-1711, 1717-1721, 1728-1732, 1738-1753, 1759-1775, 1777-1786, 1790-1794, 1796-1804,1810-1814, 1820-1829, 1831-1843, 1847-1850, 1856, 1861-1885, 1891-1900, 1906-1915, 1921-1925, 1931-1935, 1941-1950, 1975, 1986-2015, 2017-2022, 2024-2027, 2039, 2041-2044, 2046-2058, 2064, 2068, 2069, 2078-2085, 2087-2096, 2108, 2116, 2118-2122, 2132-2137, 2143- 2148, 2156-2176, 2183, 2184, 2188, 2191-2194, 2200-2213, 2215, 2220-2223, 2225-2227, 2232-2242, 2247, 2248, 2254, 2259, 2261, 2262, 2264, 2269-2272, 2278, 2280-2284, 2292-2294, 2299-2312, 2330-2337, 2353, 2359-2363, 2376, 2378-2388, 2398-2402, 2409-2416, 2427-2431, 2437, 2438, 2445-2452, 2459-2462, 2468, 2469, 2473-2475, 2484-2486, 2505-2513, 2518-2520, 2526-2540, 2557-2564, 2583-2586, 2604-2628, 2643-2645, 2648-2655, 2657-2660, 2663, 2681-2688, 2692-2694, 2697-2699, 2709, 2717-2721, 2728-2736, 2738-2741, 2748-2756, 2774-2782, 2785-2787, 2789-2792, 2796-2799, 2803, 2809-2814, 2821-2836, 2840-2845, 2850-2859, 2861-2864, 2871-2879, 2896-2900, 2903-2905, 2909-2912, 2916-2919, 2922, 2923, 2927-2936, 2941, 2948-2951, 2953-2957, 2976-2978, 2984-2989, 3000-3007, 3014-3021, 3028-3039, 3042-3051, 3056-3072, 3076-3082, 3088-3095, 3097-3114, 3117-3124, 3126-3130, 3136-3138, 3147-3151, 3154, 3163-3168, 3183-3189, 3197-3213, 3221-3224, 3229-3231, 3233-3237, 3246, 3247, 3250-3260, 3269-3286, 3290, 3307-3313, 3324-3328, 3335-3339, 3350-3353, 3358-3368, 3374-3382, 3392, 3393, 3398-3401, 3407-3409, 3411-3415, 3428-3433, 3438-3440, 3448, 3457, 3463-3468, 3473-3479, 3483-3488, 3491-3499, 3504-3508, 3511-3517, 3519-3533, 3536-3540, 3546-3570, 3574-3589, 3610, 3611, 3620-3623, 3632-3635, 3640-3642, 3649-3660, 3662-3668, 3682-3697, 3704-3714, 3717-3721, 3727-3737, 3745-3752, 3758-3768, 3781-3783, 3788-3798, 3805-3810, 3817-3826, 3831-3837, 3842-3851, 3882-3891, 3900, 3901, 3907-3911, 3922, 3923, 3932-3936, 3953-3961, 3966-3970, 3975-3985, 3990-3994, 3999-4005, 4030-4037, 4044-4049, 4056-4069, 4076-4081, 4090-4095, 4100-4104, 4109-4113, 4133-4142, 4149-4168, 4186-4202, 4213-4218, 4225-4229, 4242-4259, 4281-4289, 4300-4306, 4311-4317, 4324-4334, 4345-4350, 4353-4371, 5011-5035, 5040-5047, 5057-5064, 5078-5081, 5086-5093, 5102-5117, 5126-5141, 5150-5157, 5170-5177, 5180-5189, 5191-5196, 5199-5209, 5214-5221, 5226-5237, 5246-5249, 5254-5273, 5278-5281, 5286-5289, 5294-5297, 5306-5317, 5322-5329, 5345-5348, 5350-5353, 5358-5369, 5375-5382, 5387, 5388, 5500-5515, 5527-5531, 5542-5556, 5562-5564, 5566, 5569, 5571, 5572, 5574-5577, 5581-5583, and 5593-5600.