Z-1380-2012

CDI Blood Parameter Monitoring System 500, 500AV. Intended for use during cardiopulmonary bypass procedures

Terumo Cardiovascular Systems Corporation / initiated 2012-02-03 / Terminated / root cause: labeling and UDI

Terumo Cardiovascular Systems Corporation began a recall of CDI Blood Parameter Monitoring System 500, 500AV. Intended for use during cardiopulmonary bypass procedures on . The reason FDA records is "Replacement of the SBC batteries (system batteries) should be managed as "routine maintenance" instead of as an in-house Service Procedure because some systems may be relying on batteries beyond their useful life". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1380-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
CDI Blood Parameter Monitoring System 500, 500AV. Intended for use during cardiopulmonary bypass procedures.
Product code
DRY Monitor, Blood-Gas, On-Line, Cardiopulmonary Bypass
Reason for recall
Replacement of the SBC batteries (system batteries) should be managed as "routine maintenance" instead of as an in-house Service Procedure because some systems may be relying on batteries beyond their useful life.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
On 2/3/12 all consignees received an URGENT MEDICAL DEVICE RECALL letter via Federal Express. The letter described the reason for the recall, identified the affected product population, what to do in the event of a failure, the potential hazard, and correction. Terumo CVS will update the Operator's Manual for the CDI 500 with instructions to replace the battery within its useful life. Users are to review the notice and ensure that all users are made aware of the issue. In addition, the Customer Response Form should be completed and returned. Users can contact their local Terumo Cardiovascular Systems representative or call 1-800-521-2818 with questions or concerns, to check the last time system batteries were replaced, or to schedule maintenance for systems approaching 10 years.
Quantity in commerce
3,829 units
Distribution
Worldwide Distribution -- USA, including the states of AK, AL, AR, AZ, CA, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, WV, WA, and WI and the countries of Japan, Germany, Barbados, Saudi Arabia, Columbia, Uruguay, Ecuador, Trinidad, Costa Rica, Paraguay, Panama, Belgium, Jordan, Libya, Mexico, Brazil, Honduras, Canada, Korea, Dominica Republic, El Salvador, Guatemala, Vietnam, Chile, Hong Kong, United Arab Emirates, Singapore, Australia, Malaysia, Taiwan, Philippines, Mexico, Thailand, Vietnam, and South Africa.
Product codes and lots
500AV CDI Blood Parameter Monitoring System 500 with Arterial and Venous blood parameter modules: 1192, 1193, 1355, 1356, 1420, 1423, 1464, 1465, 1467, 1468, 1483, 1504, 1521-1524, 1526, 1527, 1529, 1530, 1602-1604, 1606, 1662, 2023, 2035, 2102-2107, 2800, 2901, 2902, 3131-3135, 3261, 3262, 4132, 5330-5333, 5338-5340, and 5579.