Smith & Nephew, SPIDER2 Limb Positioner, including accessories Battery, foot pedal and charger Catalog...
Smith & Nephew, Inc. Endoscopy Division began a recall of Smith & Nephew, SPIDER2 Limb Positioner, including accessories Battery, foot pedal and charger Catalog Number: 72203299. Table, Operating-room, Pneumatic on . The reason FDA records is "SPIDER2 may unlock (loss of traction) resulting in potential movement of the patient when the device is within close proximity to an RF source. Loss of traction can potentially result in the unintended repositioning of...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0519-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- Smith & Nephew, SPIDER2 Limb Positioner, including accessories Battery, foot pedal and charger Catalog Number: 72203299. Table, Operating-room, Pneumatic.
- Product code
- FQO Table, Operating-Room, Ac-Powered
- Reason for recall
- SPIDER2 may unlock (loss of traction) resulting in potential movement of the patient when the device is within close proximity to an RF source. Loss of traction can potentially result in the unintended repositioning of the extremity and potential for tissue or cartilage damage
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Smith & Nephew Advanced Surgical Devices Division sent an "URGENT- PRODUCT RECALL NOTIFICATION-URGENT" letter dated July 26, 2011 to all affected customers. The letter described the product, problem, and the actions to be taken by the customers. Customers were instructed to complete and return a form via fax at 405-917-8650. Contact the firm at 405-917-8556 for questions regarding this recall.
- Quantity in commerce
- 52 units
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of Canada and France.
- Product codes and lots
- Serial Numbers: VAL07, VAL08,VAL11,VAL15, VAL16, VAL19, VAL20, 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 22, 23, 24, 25, 26, 27, 28, 30, 31, 32, 33, 34, 35, 36, 37, 38, 39, 40, 41, 42, 43, 44, 45, 46, 47, 48, 49, 50, 51, 52, 53, 54, 55, 56, 58, 59, 60, 61, 62, 63, 66, 71, 73, 76, 77, 79, 80, 81, 82, 88, 90, 92, 93, 94, 95, 97, 98, 99, 100, 101, 102, 103, 104, 106, 107, 110