Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0747-2012

Olympus Electrosurgical Unit. The Olympus Electrosurgical Unit (ESG-100) is used in conjunction with cutting...

OLYMPUS Corporation of the Americas / initiated 2011-11-28 / Terminated / root cause: manufacturing process

Olympus America Inc. began a recall of Olympus Electrosurgical Unit. The Olympus Electrosurgical Unit (ESG-100) is used in conjunction with cutting and coagulation electrosurgical devices intended... on . The reason FDA records is "Investigations of customer complaints revealed a counterfeit capacitor was used in the manufacture of some lots of the ESG-100". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0747-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
OLYMPUS Corporation of the Americas
Product
Olympus Electrosurgical Unit. The Olympus Electrosurgical Unit (ESG-100) is used in conjunction with cutting and coagulation electrosurgical devices intended for use in electrosurgery procedures during flexible gastrointestinal endoscopy.
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
Investigations of customer complaints revealed a counterfeit capacitor was used in the manufacture of some lots of the ESG-100.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Olympus America Inc. sent an URGENT: MEDICAL DEVICE CORRECTION ACTION" letter dated November 28, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter instructs customers to discontinue use of the affected product. A questionnaire was attached to the letter for customers to complete and return via fax at 484-896-7128. Contact the firm at 484-896-5000 for questions regarding this notice.
Quantity in commerce
245
Distribution
Nationwide Distribution.
Product codes and lots
Serial numbers 11053P230001, 11053P230002, 11053P230003, 11053P230004, 11053P230005, 11060P230001, 11160P230002, 11060P230003, 11160P230004, 11160P230005, 11081P230001, 11081P230003, 11081P230004, 11081P230005, 11081P230006, 11081P230007, 11081P230008, 11081P230009, 11081P230010, 11094P230006, 11094P230007, 11094P230008, 11094P230009, 11094P230010, 11096P230007, 11096P230008, 11096P230009, 11096P230010, 11102P230001, 11102P230002, 11102P230003, 11102P230004, 11102P230005, 11109P230002, 11109P230003, 11109P230004, 11109P230005, 11110P230001, 11110P230002, 11110P230003, 11110P230004, 11110P230005, 11116P230001, 11116P230002, 11116P230003, 11116P230004, 11116P230005, 11117P230001, 11117P230002, 11117P230003, 11117P230004, 11117P230005, 11132P230001, 11132P230002, 11132P230003, 11132P230004, 11132P230005, 11133P230001, 11133P230002, 11133P230003, 11138P230002, 11138P230003, 11138P230004, 11139P230001, 11139P230002, 11139P230005, 11167P230001, 11167P230004, 11180P230003, 11180P230005, 11201P230005, 11201P230010, 11202P230001, 11202P230005, 25456W23-101, 25456W23-102, 25456W23-107, 25456W23-108, 25456W23-109, 25522W23-101, 25522W23-102, 25522W23-103, 25522W23-104, 25522W23-105, 25553W23-105, 25553W23-107, 25553W23-108, 25553W23-109, 25553W23-110, 25565W23-102, 25565W23-103, 25565W23-105, 25565W23-106, 25565W23-108, 25565W23-109, 25565W23-110, 25593W23-101, 25593W23-102, 25593W23-103, 25593W23-104, 25593W23-105, 25593W23-107, 25593W23-109, 25593W23-110, 25607W23-101, 25607W23-102, 25607W23-103, 25607W23-104, 25607W23-105, 25607W23-106, 25607W23-107, 25607W23-108, 25607W23-110, 25613W23-101, 25613W23-102, 25613W23-103, 25613W23-104, 25613W23-105, 25613W23-106, 25613W23-107, 25613W23-108, 25613W23-109, 25613W23-110, 25686W23-102, 25719W23-105, 25719W23-106, 25736W23-101, 25736W23-102, 25736W23-103, 25736W23-104, 25736W23-105, 25736W23-106, 25736W23-107, 25736W23-108, 25736W23-109, 25752W23-101, 25752W23-102, 25752W23-103, 25752W23-104, 25752W23-105, 25752W23-106, 25752W23-107, 25776W23-101, 25776W23-102, 25776W23-103, 25776W23-105, 25776W23-106, 25776W23-107, 25776W23-108, 25776W23-109, 25776W23-110, 25869W23-101, 25869W23-102, 25869W23-103, 25869W23-104, 25869W23-105, 25869W23-106, 25869W23-107, 25869W23-108, 25869W23-109, 25869W23-110, 25978W23-104, 25978W23-105, 25978W23-106, 25978W23-107, 26052W23-109, 26052W23-110, 26085W23-101, 26085W23-102, 26085W23-103, 26085W23-104, 26085W23-105, 26085W23-106, 26085W23-107, 26085W23-108, 26085W23-109, 26136W23-101, 26136W23-103, 26136W23-104, 26136W23-105, 26136W23-107, 26136W23-108, 26136W23-109, 26136W23-110, 26162W23-108, 26162W23-109, 26162W23-110, 26227W23-102, 26227W23-103, 26227W23-104, 26227W23-105, 26227W23-106, 26227W23-107, 26227W23-108, 26227W23-109, 26227W23-110, 26456W23-101, 26456W23-102, 26456W23-103, 26456W23-104, 26456W23-105, 26456W23-106, 26456W23-107, 26456W23-108, 26456W23-109, 26456W23-110, 26559W23-101, 26559W23-103, 26559W23-104, 26559W23-105, 26559W23-108, 26559W23-109, 26559W23-110, 26710W23-108, 26710W23-109, 26770W23-101, 26770W23-102, 26770W23-103, 26770W23-104, 26770W23-105, 26770W23-106, 26770W23-107, 26770W23-108, 26770W23-109, 26770W23-110, 26878W23-101, 26878W23-102, 26878W23-103, 26878W23-104, 26878W23-105, 26879W23-102, 26879W23-103, 26879W23-104, 26879W23-105, 26917W23-101, 26917W23-102, 26917W23-103, 26917W23-105, 26917W23-106, 26917W23-107, 26917W23-108, 26917W23-109, 26917W23-110, and 26992W23-110.