Z-0598-2012
Baxter Colleague CX Infusion Pump, Single Channel Volumetric Infusion Pumps
Baxter Healthcare Corporation / initiated 2009-05-08 / Terminated / root cause: manufacturing process
Baxter Healthcare Corp. began a recall of Baxter Colleague CX Infusion Pump, Single Channel Volumetric Infusion Pumps. Product Code: 2M8161 on . The reason FDA records is "There was no documentation to show the required tests were performed following the replacement of the battery and battery harness". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0598-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Baxter Colleague CX Infusion Pump, Single Channel Volumetric Infusion Pumps. Product Code: 2M8161.
- Product code
- FRN Pump, Infusion
- Reason for recall
- There was no documentation to show the required tests were performed following the replacement of the battery and battery harness.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Baxter sent a service technician to the consignee location on 05/08/2009 to perform the battery check and discharge test. The pump was found to be within specification.
- Quantity in commerce
- one pump
- Distribution
- Puerto Rico
- Product codes and lots
- Product Code: 2M8161, Serial Number 12050019CC