Z-0598-2012

Baxter Colleague CX Infusion Pump, Single Channel Volumetric Infusion Pumps

Baxter Healthcare Corporation / initiated 2009-05-08 / Terminated / root cause: manufacturing process

Baxter Healthcare Corp. began a recall of Baxter Colleague CX Infusion Pump, Single Channel Volumetric Infusion Pumps. Product Code: 2M8161 on . The reason FDA records is "There was no documentation to show the required tests were performed following the replacement of the battery and battery harness". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0598-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
Baxter Colleague CX Infusion Pump, Single Channel Volumetric Infusion Pumps. Product Code: 2M8161.
Product code
FRN Pump, Infusion
Reason for recall
There was no documentation to show the required tests were performed following the replacement of the battery and battery harness.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Baxter sent a service technician to the consignee location on 05/08/2009 to perform the battery check and discharge test. The pump was found to be within specification.
Quantity in commerce
one pump
Distribution
Puerto Rico
Product codes and lots
Product Code: 2M8161, Serial Number 12050019CC