Z-0445-2012

Crystalens Accommodating Posterior Chamber Intraocular Lens, AT-50AO. Crystalens Accommodating Posterior...

Bausch & Lomb Incorporated / initiated 2011-11-14 / Terminated / root cause: manufacturing process

Bausch and Lomb, Incorporated began a recall of Crystalens Accommodating Posterior Chamber Intraocular Lens, AT-50AO. Crystalens Accommodating Posterior Chamber Intraocular Lens PMA No.: P030002. The... on . The reason FDA records is "Bausch + Lomb voluntary initiated the recall due to a physician complaint of haze on the Crystalens AO which was visualized using an angled beam from a slit lamp during the post-operative examination". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0445-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bausch & Lomb Incorporated
Product
Crystalens Accommodating Posterior Chamber Intraocular Lens, AT-50AO. Crystalens Accommodating Posterior Chamber Intraocular Lens PMA No.: P030002. The Crystalens is intended for primary implantation in the capsular bag of the eye for the visual correction of aphakia secondary to the removal of a cataractous lens in adult patients with and without presbyopia.
Product code
NAA Lens, Intraocular, Accommodative
Reason for recall
Bausch + Lomb voluntary initiated the recall due to a physician complaint of haze on the Crystalens AO which was visualized using an angled beam from a slit lamp during the post-operative examination.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Bausch + Lomb sent an "URGENT- MEDICAL DEVICE RECALL" letter dated November 14, 2011 to all affected customers. The letter described the product, problem, and actions to be taken by the customers. The letter instructs customers to remove the affected products from inventory and return the lenses to the firm. Additionally, a Recall Acknowledgement Form was attached to the letter for customers to complete and return with the affected products. Contact Bausch + Lomb at 949-521-7895 for questions regarding this recall..
Quantity in commerce
62 units
Distribution
Nationwide Distribution
Product codes and lots
Serial No.: Expiration Date: 11-699865 to 11-699951 01/14/2016 11-699952 to 11-700027 01/14/2016