Z-0464-2012

CD HORIZON (R) Spinal System, MULTI-AXIAL SCREW FOR 5.5mm ROD W/DUAL LEAD THREAD, Size 6.5mm X 40mm, Mat'l...

Medtronic Sofamor Danek USA Inc / initiated 2011-11-30 / Terminated

Medtronic Sofamor Danek USA Inc began a recall of CD HORIZON (R) Spinal System, MULTI-AXIAL SCREW FOR 5.5mm ROD W/DUAL LEAD THREAD, Size 6.5mm X 40mm, Mat'l TITANIUM ALLOY, REF X0809439, Rx only, Medtronic... on . The reason FDA records is "One screw may be longer then size etched on the screw". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0464-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Sofamor Danek USA Inc
Product
CD HORIZON (R) Spinal System, MULTI-AXIAL SCREW FOR 5.5mm ROD W/DUAL LEAD THREAD, Size 6.5mm X 40mm, Mat'l TITANIUM ALLOY, REF X0809439, Rx only, Medtronic Sofamor Danek USA, 1800 Pyramid Place, Memphis, TN 38132 implantation
Product code
KWQ Appliance, Fixation, Spinal Intervertebral Body
Reason for recall
One screw may be longer then size etched on the screw.
FDA root cause
Mixed-up of materials/components
Action
Medtronic Sofamor Danek USA, Inc sent a Voluntary Field Correction letter dated November 30, 2011, via FedEx to all affected customers. The letter identified the product, the problem and the action needed to be taken by the customer. The firm will visit each consignee to perform a visual check of inventory. Please utilize the attached questionnare to account for the affected product and to record any patient injuries that may have occurred specifically related to the use of these items. If you need assistance, or if you have any related questions or concerns please call (800) 876-3133 extension 3173
Quantity in commerce
252 units
Distribution
Nationwide Distribution including the states of AL, PA, IN, CO, MI and IA.
Product codes and lots
Lot Numbers: 0132008W, 0150703W, 0150705W, and 0166823W