Biomet Series A Thin Patella size 28mm, three pegs - 184782, Sterile, Biomet Orthopedics, Warsaw, IN 46582...

Zimmer GmbH / initiated 2011-10-04 / Terminated / root cause: labeling and UDI

Biomet, Inc. began a recall of Biomet Series A Thin Patella size 28mm, three pegs - 184782, Sterile, Biomet Orthopedics, Warsaw, IN 46582 For use with Vanguard, Maxim, Ascent & AGC Knee... on . The reason FDA records is "Correcting language on product labeling which incorrectly states patella buttons are intended for use with the AGC knee system". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0702-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Biomet Series A Thin Patella size 28mm, three pegs - 184782, Sterile, Biomet Orthopedics, Warsaw, IN 46582 For use with Vanguard, Maxim, Ascent & AGC Knee Systems Knee joint patellofemorotibial, cemented, polymer/metal/polymer, semi-constrained. Knee joint replacement prosthesis, intended for patella replacement of the knee joint in conjunction with a femoral and tibial component.
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
Correcting language on product labeling which incorrectly states patella buttons are intended for use with the AGC knee system.
Root cause tag
labeling and UDI
FDA root cause
Error in labeling
Action
Biomet sent a "FIELD CORRECTION NOTICE" dated October 3, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter instructs customers to locate the affected product and to follow the work instructions included in the letter. Additionally, a Confirmation of Completion Response Form was enclosed for customers to complete and return via fax to 574-372-1683. Questions related to this notice are directed to 574-371-3755.
Distribution
Worldwide Distribution-USA (nationwide) and Puerto Rico and the countries of Argentina, Australia, Canada, Chile, Costa Rica, Europe, India, Japan, Korea, and Singapore.
Product codes and lots
Part 184782: Lot 098970, 144970, 190440, 232540, 274550, 357320, 397810, 397820, 438790, 438800, 498310, 539710, 599220, 644430, 710240