Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0745-2012

Baxter Colleague 3 Triple Channel Volumetric Infusion Pump. Baxter Healthcare Corporation, Medication...

Baxter Healthcare Corporation / initiated 2010-07-14 / Terminated / root cause: manufacturing process

Baxter Healthcare Corp. began a recall of Baxter Colleague 3 Triple Channel Volumetric Infusion Pump. Baxter Healthcare Corporation, Medication Delivery Division, Deerfield, IL 60015. Product Code... on . The reason FDA records is "Baxter's Service Center did not perform the ground impedance and leakage current test which verify if any power overflow occurs and that it is properly discharged through the device electrical system". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0745-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
Baxter Colleague 3 Triple Channel Volumetric Infusion Pump. Baxter Healthcare Corporation, Medication Delivery Division, Deerfield, IL 60015. Product Code: 2M8153. Product Usage: Intended for use in a wide variety of patient environments for adult, pediatric and neonatal patients. Electronic infusion pumps indicated for continuous or intermittent delivery of solution through clinically acceptable routes of administration, such as intravenous (IV), intra-arterial (IA), subcutaneous, epidural or irrigation of fluid spaces applications.
Product code
FRN Pump, Infusion
Reason for recall
Baxter's Service Center did not perform the ground impedance and leakage current test which verify if any power overflow occurs and that it is properly discharged through the device electrical system.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Baxter sent service technicians to the consignees location on July 14, 2010 and October 6, 2010 and conducted the ground impedance and leakage current test on the affected pumps. The pumps were found with specifications.
Quantity in commerce
1 unit
Distribution
Nationwide distribution: Puerto Rico.
Product codes and lots
Product Code: 2M8153, Serial Number 14030127CT