Agfa Computed Radiography Systems with DX-G Digitizer. DX-G Digitizers haved the same intended use as the...
AGFA Corp. began a recall of Agfa Computed Radiography Systems with DX-G Digitizer. DX-G Digitizers haved the same intended use as the predicate devices. they are intended to use in... on . The reason FDA records is "Occasionally images can be lost or sent to an incorrect folder or the digitizer may stop either during startup or during a cassette scan cycle". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0591-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- AGFA Healthcare Corp.
- Product
- Agfa Computed Radiography Systems with DX-G Digitizer. DX-G Digitizers haved the same intended use as the predicate devices. they are intended to use in providing diagnostic images to aid the physician with diagnosis. DX-G Digitizers provide the ability to use both standand needle phosphor image plates and can be used with either Musica or Musica 2. In the US, Agfa's Computed Radiography Systems with DX-G Digitizers are not intended for use in mammography.
- Product code
- MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)
- Reason for recall
- Occasionally images can be lost or sent to an incorrect folder or the digitizer may stop either during startup or during a cassette scan cycle.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Agfa HealthCare sent a 'URGENT SAFETY NOTICE" letter dated December 6, 2011, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Agfa Service will contact your facility and schedule a software upgrade if this has not already been performed. Please complete the attached feedback form as soon as possible and fax it to 864-421-1662. If you have any questions about this matter, please contact your local Agfa HealthCare Service at 877-777-2432.
- Quantity in commerce
- 125
- Distribution
- Worldwide Distribution-- USA (nationwide) including the states of AZ, CA, CO, FL, GA, HI, IL, IN, KS, LA, ME, MI, MN, MO, NE, NH, NJ, NY, OH, OK, PA, SD, TX, VA, WA, WI and WV, and Canada.
- Product codes and lots
- Model No. 5170/100