ThermoSuit Manufactured for Life Recovery Systems. HD, LLC, 170 Kinnelon Rd. Kinnelon, NJ 07405 USA Tel...
Life Recovery Systems HD, LLC began a recall of ThermoSuit Manufactured for Life Recovery Systems. HD, LLC, 170 Kinnelon Rd. Kinnelon, NJ 07405 USA Tel.: (973) 283-2800 FAx.: (973) 283-2910 Temperature... on . The reason FDA records is "ThermoSuit System Model TSP-50 Pump/Controller may have corrosion on the temperature circuit board which may result in incorrect patient temperature reading". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0456-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Life Recovery Systems HD, LLC
- Product
- ThermoSuit Manufactured for Life Recovery Systems. HD, LLC, 170 Kinnelon Rd. Kinnelon, NJ 07405 USA Tel.: (973) 283-2800 FAx.: (973) 283-2910 Temperature reduction in patients where clinically indicated; Monitoring of patient temperature
- Product code
- NZE Hyperthermia Monitor
- Reason for recall
- ThermoSuit System Model TSP-50 Pump/Controller may have corrosion on the temperature circuit board which may result in incorrect patient temperature reading.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Life Recovery Systems sent recall letters/return response forms dated 10/24/2011 to all consignees. Life Recovery Systems has expanded their recall to include additional units and to provide additional service to products out in the field. A new recall letter was sent to consignees on 4/27/2012.
- Quantity in commerce
- 28 units
- Distribution
- Nationwide Distribution including NY, NC, LA, NJ, AL. PA TX, KY, MO, MA, and TN
- Product codes and lots
- 510 K K061023 N0110PD N0108NI N0068JH N0109PD N0103NE N0106NI N0065JH N0104NI N0105NI