Siemens CLINITEST hCG -Pregnancy Test (qualitative), Radio immunoassay,Human Chorionic Gonadotropin (hCG)...
Siemens Healthcare Diagnostics began a recall of Siemens CLINITEST hCG -Pregnancy Test (qualitative), Radio immunoassay,Human Chorionic Gonadotropin (hCG) test system. Package of 25. Product Code 1760/... on . The reason FDA records is "A potential for false positive or increase in borderline results with patients and commercially available negative control material". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0972-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Siemens CLINITEST hCG -Pregnancy Test (qualitative), Radio immunoassay,Human Chorionic Gonadotropin (hCG) test system. Package of 25. Product Code 1760/ Catalog# 06484105). For in vitro diagnostic use as a qualitative method in the rapid detection of human chorionic gonadotropin (hCG) in urine specimens.
- Product code
- JHI Visual, Pregnancy Hcg, Prescription Use
- Reason for recall
- A potential for false positive or increase in borderline results with patients and commercially available negative control material.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Siemens issued Overnight Mailings of Urgent: Recall Notices on December 15, 2011 for delivery beginning December 16, 2011 to Distributors and end users. Customer bulletin has been issued instructing customers to discontinue use of product and discard impacted inventory. Customers are to contact their distributor regarding credit for unused product. For technical questions contact Siemens Technical Solutions Center at 877-229-3711.
- Quantity in commerce
- 48,578 Kits (25 tests / kit)
- Distribution
- Worldwide Distribution -- US, Austria, Germany, Great Britain, Denmark, Norway, Sweden, Poland, Netherlands, Luxembourg, Belgium, Finland, France, Greece, Ireland, Italy, Portugal, Qatar, Saudi Arabia, Slovakia, Slovenia, and Switzerland.
- Product codes and lots
- Lot Expiration 028619, 6/8/2012; 028924, 7/5/2012; 029997, 10/7/2012; 030240, 11/6/2012; 030431, 11/24/2012; 030602, 12/8/2012; 030826, 12/24/2012; 030998, 1/11/2013; 031332, 2/4/2013; 031460, 2/19/2013; 031662, 3/6/2013; 031798, 3/20/2013; 031924, 3/30/2013; 032180, 4/28/2013; and 032106, 4/19/2013.