Z-0689-2012

MAQUET QUADROX iD PEDIATRIC -OXYGENATOR: BEQ-HMOD 3000; Made In Germany The diffusion membrane oxygenator...

Maquet Cardiovascular Us Sales, LLC / initiated 2011-09-22 / Terminated / root cause: design

Maquet Cardiovascular Us Sales, Llc began a recall of MAQUET QUADROX iD PEDIATRIC -OXYGENATOR: BEQ-HMOD 3000; Made In Germany The diffusion membrane oxygenator Quadrox iD pediatric is intended for use in an... on . The reason FDA records is "There are reports of leakage in the Leuer connector of the blood outflow connector of the Quadrox-iD pediatric oygenator". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0689-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Maquet Cardiovascular Us Sales, LLC
Product
MAQUET QUADROX iD PEDIATRIC -OXYGENATOR: BEQ-HMOD 3000; Made In Germany The diffusion membrane oxygenator Quadrox iD pediatric is intended for use in an extracorporeal perfusion system. The oxygenator is designed for a blood flow rate of 0.2 -2.8 I/min and is intended for the treatment of pediatric patients. Within the specified flow rate range, the device oxygenates the blood, removes carbon dioxide and regulates the blood temperature.The utilization period for this device is restricted to six hours.
Product code
DTZ Oxygenator, Cardiopulmonary Bypass
Reason for recall
There are reports of leakage in the Leuer connector of the blood outflow connector of the Quadrox-iD pediatric oygenator.
Root cause tag
design
FDA root cause
Device Design
Action
The firm, Maquet Getinge Group, sent an "PRODUCT RECALL URGENT-MEDICAL DEVICE REMOVAL ACTION" letter dated September 22, 2011 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed tocheck their remaining inventory; remove product immediately to prevent use; remove the boxes or individually pouched oxygenators in a secure area of their facility and notify their MAQUET representative (a representative will physically remove the recalled product and replenish the removed units); and complete and return the enclosed Product Removal Disposition Form via fax to: (973)-807-9210 or email: [email protected]. Should you have any questions or concerns, please do not hesitate to contact MAQUET Customer Service at 800-777-4222.
Quantity in commerce
168 units
Distribution
Nationwide distribution: AL, CA, DE, DC, FL, GA, HI, IL, IN, LA, MI, MO, MS, NC, NE, NJ, NM, NY, OH, OK, PA, and TX.
Product codes and lots
510(k) numbers: 1<100278 (Bioline coated) Device Listing No. 0098004 BEQ-HMOD 30000 70064002-28 units 70064004-140 units