Z-0714-2012

VITROS Chemistry Products Calibrator Kit 16, REF 680 1700, 5x1 mL --- Ortho-Clinical Diagnostics, Rochester...

Ortho-Clinical Diagnostics, Inc. / initiated 2011-11-10 / Terminated / root cause: manufacturing process

Ortho-Clinical Diagnostics began a recall of VITROS Chemistry Products Calibrator Kit 16, REF 680 1700, 5x1 mL --- Ortho-Clinical Diagnostics, Rochester, NY USA -- Ortho-Clinical Diagnostics High... on . The reason FDA records is "Ortho Clinical Diagnostics is recalling Lot 1630 of VITROS Calibrator Kit 16 because it will not likely meet its specified 12 months of stability". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0714-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS Chemistry Products Calibrator Kit 16, REF 680 1700, 5x1 mL --- Ortho-Clinical Diagnostics, Rochester, NY USA -- Ortho-Clinical Diagnostics High Wycomber, UK. VITROS Chemistry Products Calibrator Kit 16 is used in conjunction with VITROS Chemistry Products FS Calibrator 1 to calibrate the VITROS 5,1 FS Chemistry System, the VITROS 4600 Chemistry System and the VITROS 5600 Integrated System for the quantitative measurement of rheumatoid factor (RF).
Product code
DHR System, Test, Rheumatoid Factor
Reason for recall
Ortho Clinical Diagnostics is recalling Lot 1630 of VITROS Calibrator Kit 16 because it will not likely meet its specified 12 months of stability.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
A letter was sent to customers via overnight delivery on 11/10/2011. The letter informed customers they should discontinue using and discard all remaining VITROS Calibrator Kit 16, Lot 1630; Complete and return the attached Confirmation of Receipt form upon receipt of the notification. Once the fax reply is received, the firm will begin the replacement order process; forward the notification to all departments and areas that use the VITROS Calibrator Kit 16, Lot 1630; and post the notification by each VITROS Chemistry or Integrated System in your facility that uses the VITROS Calibrator Kit 16, Lot 1630 or with VITROS user documentation.
Quantity in commerce
768 kits
Distribution
Class II Recall: Worldwide Distribution - USA (Nationwide) and the countries of Brazil, Canada, Chile, China, Colombia, England, France, Mexico, Panama, Singapore, Spain, Venezuela.
Product codes and lots
Lot 1630