Z-0802-2013Class II

Terumo CDI 101 Hematocrit/Oxygen Saturation Monitoring System. Monitors blood/oxygen saturation, hematocrit...

Terumo Cardiovascular Systems Corporation / initiated 2010-11-01 / Terminated / root cause: software

Terumo Cardiovascular Systems Corporation began a recall of Terumo CDI 101 Hematocrit/Oxygen Saturation Monitoring System. Monitors blood/oxygen saturation, hematocrit and hemoglobin levels on . FDA classified it as Class II and gives the reason as "Inaccurate readings. Software contains unused (turned off) accuracy correction factors for slope (gain) and offset (bias) for Hematocrit/Saturation calculations. The accuracy on any cuvette size did not meet the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0802-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Terumo CDI 101 Hematocrit/Oxygen Saturation Monitoring System. Monitors blood/oxygen saturation, hematocrit and hemoglobin levels.
Product code
DRY Monitor, Blood-Gas, On-Line, Cardiopulmonary Bypass
Reason for recall
Inaccurate readings. Software contains unused (turned off) accuracy correction factors for slope (gain) and offset (bias) for Hematocrit/Saturation calculations. The accuracy on any cuvette size did not meet the specifications.
Root cause tag
software
FDA root cause
Software design
Action
Terumo sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated November 1, 2010 to all affected customers. The letter identified the product, problem, and the actions to be taken by the customers. Contact Terumo CVS Customer Service at 1-800-521-2818 for questions or concerns.
Quantity in commerce
622 units
Distribution
Worldwide Distribution.
Product codes and lots
Catalog number: 101 and Serial Numbers: 0503, 1000, 1017, 1070, 1071, 1073, 1074, 1076, 1077, 1080, 1082, 1083, 1107-1110, 1115-1117,1124,1126-1134, 1136, 1137, 1139-1146, 1148-1152, 1154-1156, 1159-1162, 1164-1166, 1168-1177, 1182, 1185-1190, 1192,1193, 1195, 1197, 1198, 1225, 1228, 1229, 1237, 1239-1241,1243-1245,1247, 1249-1326, 1332-1339, 1344-1347, 1349, 1350-1363, 1365-1409, 1411-1436,1438, 1440-1447, 1449-1453, 1455-1463,1465-1482,1484-1492,1494-1552,1555-1565, 1567-1600,1602,1605-1607,1609-1611,1613-1623,1625-1629,1632-1651,1653, 1655-1660, 1662-1664, 1666, 1667,1669-1679,1681-1684,1686,1689-1691, 1693-1710,1712-1715,1717-1750,1752-1757,1759-1771,1773-1785,1787,1788,1790-1809, 1811-1831,1833,1838,1839,1855-1858,1866, and TK001-TK011.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0802-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0802-2013Class IITerminatedVoluntary: Firm initiated