Micro-Tech Informer Plus Silver Monitor, Model 81830. The firm name on the label is Stanley Security...
Stanley Security Solutions, Inc. began a recall of Micro-Tech Informer Plus Silver Monitor, Model 81830. The firm name on the label is Stanley Security Solutions, Inc., Lincoln, NE. Product Usage: The system is... on . The reason FDA records is "Connection between the call cord and the receptacle may not make a correct connection resulting in a continuous or intermittent (false) signal to the nurse call system". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1476-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Securitas Healthcare LLC
- Product
- Micro-Tech Informer Plus Silver Monitor, Model 81830. The firm name on the label is Stanley Security Solutions, Inc., Lincoln, NE. Product Usage: The system is a bed exit system designed to indicate, by alarm or signal, when someone attempts to leave the bed. The system consists of a control unit that is mounted on the bed and a pressure sensitive bed sensor pad that is placed across the width of the bed under the mattress or just beneath the sheets and overlays. The control unit is powered by means of a plug-in type power supply and an alarm cord is used to interface the control unit with the nurse call system.
- Product code
- KMI Monitor, Bed Patient
- Reason for recall
- Connection between the call cord and the receptacle may not make a correct connection resulting in a continuous or intermittent (false) signal to the nurse call system
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Stanley Healthcare Solutions, Inc. contacted affected customers via phone and email on December 13, 2011. A follow-up certified letter was sent on December 14, 2011. Customers were additionally sent and "Important Recall Notice letter on December 16, 2011 after errors in the original letters were detected. The letter identified the affected product, problem, actions to be taken, and included a reply form for the customer to complete and return. The recalling firm requested their customers to stop using the monitor and contact them for return of the unit. Contact information (phone, fax, & email) was included. For question contact Micro-Tech Customer Service at 1-800-206-1045.
- Quantity in commerce
- 824 units
- Distribution
- Worldwide Distribution - USA Nationwide including the states of CT, FL, KY, MN, NC, NE, NH, NJ, NY, PA, TX and the country of Canada (5 units).
- Product codes and lots
- Serial numbers 071100025 through 111100306