Z-0769-2012

SYNERGY Perfusion Tubing Systems US Denver PHISIO, Sorin Group USA, Tubing set heart-lung bypass, Sterile EO...

LivaNova USA, Inc. / initiated 2011-12-02 / Terminated / root cause: design

Sorin Group USA, Inc. began a recall of SYNERGY Perfusion Tubing Systems US Denver PHISIO, Sorin Group USA, Tubing set heart-lung bypass, Sterile EO, manufactured by: Sorin Group Italia, Via Statale... on . The reason FDA records is "Customer complaints of a leak between the blood and water compartments of the heat exchanger used in the oxygenators due to prolonged contact of the heat exchangers with sodium chloride causing micro-holes". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0769-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
LivaNova USA, Inc.
Product
SYNERGY Perfusion Tubing Systems US Denver PHISIO, Sorin Group USA, Tubing set heart-lung bypass, Sterile EO, manufactured by: Sorin Group Italia, Via Statale 12 Nord, 86, 41037 Mirandola (MO) Italy. Intended for use in small-adult or adult surgical procedures requiring extracorporeal gas exchange support and blood temperature control for periods up to 6 hours.
Product code
KFM Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type
Reason for recall
Customer complaints of a leak between the blood and water compartments of the heat exchanger used in the oxygenators due to prolonged contact of the heat exchangers with sodium chloride causing micro-holes.
Root cause tag
design
FDA root cause
Device Design
Action
Urgent Field Safety Notices dated December 2, 2011 were sent to customers. The letter identified the affected products and described the problem. The letter also provided additional directions to the Instructions for Use in order to minimize risk and clarify priming procedures. Customers were to communicate the new warning to all personnel in the use of the affected products and complete the Customer Response Form. The form should be returned via fax or e-mail. Questions should be directed towards Sorin Group Customer Service at 1-800-650-2623.
Quantity in commerce
3,310 units (sold as standalone devices)
Distribution
Worldwide Distribution
Product codes and lots
All Lots.