Z-0770-2012

Synthesis PHisio Oxygenator extracorporeal membrane, Sterile EO, Distributed in U.S. by: Sorin Group USA...

LivaNova USA, Inc. / initiated 2011-12-02 / Terminated / root cause: design

Sorin Group USA, Inc. began a recall of Synthesis PHisio Oxygenator extracorporeal membrane, Sterile EO, Distributed in U.S. by: Sorin Group USA, Inc, 14401 W. 65th Way, Arvada, CO 80004... on . The reason FDA records is "Customer complaints of a leak between the blood and water compartments of the heat exchanger used in the oxygenators due to prolonged contact of the heat exchangers with sodium chloride causing micro-holes". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0770-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
LivaNova USA, Inc.
Product
Synthesis PHisio Oxygenator extracorporeal membrane, Sterile EO, Distributed in U.S. by: Sorin Group USA, Inc, 14401 W. 65th Way, Arvada, CO 80004, manufactured by: Sorin Group Italia, Via Statale 12 Nord, 86, 41037 Mirandola (MO) Italy. Intended for use in small-adult or adult surgical procedures requiring extracorporeal gas exchange support and blood temperature control for periods up to 6 hours.
Product code
DTZ Oxygenator, Cardiopulmonary Bypass
Reason for recall
Customer complaints of a leak between the blood and water compartments of the heat exchanger used in the oxygenators due to prolonged contact of the heat exchangers with sodium chloride causing micro-holes.
Root cause tag
design
FDA root cause
Device Design
Action
Urgent Field Safety Notices dated December 2, 2011 were sent to customers. The letter identified the affected products and described the problem. The letter also provided additional directions to the Instructions for Use in order to minimize risk and clarify priming procedures. Customers were to communicate the new warning to all personnel in the use of the affected products and complete the Customer Response Form. The form should be returned via fax or e-mail. Questions should be directed towards Sorin Group Customer Service at 1-800-650-2623.
Quantity in commerce
41,201 units (11,182 as standalone devices/ 30,019 included in heart and lung packs)
Distribution
Worldwide Distribution
Product codes and lots
All Lots.