Ingenuity CT (Computed Tomography) System, 510 (k) #K033326, Model #728326, Serial Numbers: 300005, 300010...
Philips Medical Systems (Cleveland) Inc began a recall of Ingenuity CT (Computed Tomography) System, 510 (k) #K033326, Model #728326, Serial Numbers: 300005, 300010, 300011, 300012, 300013, 300014, 300015, 300016... on . The reason FDA records is "Component defect. When the user is operating the control panel, he/she may inadvertently activate the wrong switch on the control panel which could, in turn, cause the patient couch to move in an unwanted direction". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0693-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Ingenuity CT (Computed Tomography) System, 510 (k) #K033326, Model #728326, Serial Numbers: 300005, 300010, 300011, 300012, 300013, 300014, 300015, 300016, 300017, 300018, 300019, 300020 and 300021. Product Usage: The Ingenuity CT is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of X-Ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- Component defect. When the user is operating the control panel, he/she may inadvertently activate the wrong switch on the control panel which could, in turn, cause the patient couch to move in an unwanted direction.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Philips sent an 'URGENT - Medical Device Correction Ingenuity CT' letter dated November 15, 2011. The letter describes the affected product, problem, hazard involved and actions to be taken by the customer and user. In this case, the customers are advised that extra care should be taken when using the 8-way Gantry Panel switch so as to avoid inadvertently activating an unintended switch resulting in unwanted (couch) movement. The letter notifies the customers that a Philips representative will be visiting each customer site and installing a new control panel through a Field Change Order. For further information or support concerning this issue contact your local Philips representative or local Philips Healthcare office. For North America and Canada contact the Customer Care Solutions Center at (1-800-722-9377, option 5; enter site ID or follow prompts).
- Quantity in commerce
- 13 system units were distributed (5 in U.S. & 8 overseas)
- Distribution
- Worldwide Distribution - USA (nationwide) in the states of: LA, NY, OH, WV and the foreign countries of: Australia, Germany, Korea, India, Malaysiaa and New Zealand.
- Product codes and lots
- The following Serial Numbers of this device are subject to recall: Serial Numbers: 300005, 300010, 300011, 300012, 300013, 300014, 300015, 300016, 300017, 300018, 300019, 300020 and 300021.