Z-0693-2012

Ingenuity CT (Computed Tomography) System, 510 (k) #K033326, Model #728326, Serial Numbers: 300005, 300010...

Philips North America LLC / initiated 2011-11-11 / Terminated / root cause: design

Philips Medical Systems (Cleveland) Inc began a recall of Ingenuity CT (Computed Tomography) System, 510 (k) #K033326, Model #728326, Serial Numbers: 300005, 300010, 300011, 300012, 300013, 300014, 300015, 300016... on . The reason FDA records is "Component defect. When the user is operating the control panel, he/she may inadvertently activate the wrong switch on the control panel which could, in turn, cause the patient couch to move in an unwanted direction". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0693-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Ingenuity CT (Computed Tomography) System, 510 (k) #K033326, Model #728326, Serial Numbers: 300005, 300010, 300011, 300012, 300013, 300014, 300015, 300016, 300017, 300018, 300019, 300020 and 300021. Product Usage: The Ingenuity CT is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of X-Ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Component defect. When the user is operating the control panel, he/she may inadvertently activate the wrong switch on the control panel which could, in turn, cause the patient couch to move in an unwanted direction.
Root cause tag
design
FDA root cause
Device Design
Action
Philips sent an 'URGENT - Medical Device Correction Ingenuity CT' letter dated November 15, 2011. The letter describes the affected product, problem, hazard involved and actions to be taken by the customer and user. In this case, the customers are advised that extra care should be taken when using the 8-way Gantry Panel switch so as to avoid inadvertently activating an unintended switch resulting in unwanted (couch) movement. The letter notifies the customers that a Philips representative will be visiting each customer site and installing a new control panel through a Field Change Order. For further information or support concerning this issue contact your local Philips representative or local Philips Healthcare office. For North America and Canada contact the Customer Care Solutions Center at (1-800-722-9377, option 5; enter site ID or follow prompts).
Quantity in commerce
13 system units were distributed (5 in U.S. & 8 overseas)
Distribution
Worldwide Distribution - USA (nationwide) in the states of: LA, NY, OH, WV and the foreign countries of: Australia, Germany, Korea, India, Malaysiaa and New Zealand.
Product codes and lots
The following Serial Numbers of this device are subject to recall: Serial Numbers: 300005, 300010, 300011, 300012, 300013, 300014, 300015, 300016, 300017, 300018, 300019, 300020 and 300021.