STERIS SYSTEM 1E Liquid Chemical Sterilant Processing System. Model P6500, 510 (k) Numbers: K090036 Cleared...
Steris Corporation Hopkins Facility began a recall of STERIS SYSTEM 1E Liquid Chemical Sterilant Processing System. Model P6500, 510 (k) Numbers: K090036 Cleared 4/05/2010; K101409 cleared 8/2/2010; K102462... on . The reason FDA records is "Software deficiency. STERIS identified that water specifications were not being met at some of their customer sites. Critical parameters for initiating cycles are: water quality, water pressure and water temperature...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0869-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- STERIS Corporation
- Product
- STERIS SYSTEM 1E Liquid Chemical Sterilant Processing System. Model P6500, 510 (k) Numbers: K090036 Cleared 4/05/2010; K101409 cleared 8/2/2010; K102462, cleared 9/2/2010; add to file 2/24/2011. The SYSTEM 1E Liquid Sterilant Processing System is intended for the liquid chemical sterilization of manually cleaned immersible and reuseable critical and semi-critical heat sensitive medical devices, including endoscopes and their accessories.
- Product code
- MED Sterilant, Medical Devices
- Reason for recall
- Software deficiency. STERIS identified that water specifications were not being met at some of their customer sites. Critical parameters for initiating cycles are: water quality, water pressure and water temperature. Failure to meet these criteriae may lead to cycle aborts by the SYSTEM 1E Liquid Chemical Sterilant Processing System. The firm intends to perform a software field correction which
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm, STERIS Corp. prepared an "URGENT VOLUNTARY FIELD CORRECTION NOTICE" letter dated December 8, 2011. The firm intends to send the letter to its customers upon STERIS' receipt of 510(k) clearance from FDA. The letter includes a description of the Product-SYSTEM Liquid Chemical Sterilant Processing System, problem and actions taken. The customers were instructed to in case of an incomplete or cancelled cycle, reprocess devices in the cycle following the directions in the Operator manual. The letter also states that STERIS Corporation will install updated software on the affected system, and will be providing the customers with a new Operator Manual reflecting the revised wording resulting from the installation of the new software. For further information or if they have any questions regarding the firms visit to their facility, contact STERIS Field Service Dispatch at 1-800-333-8828. If you have questions regarding this matter, contact Director Low Temperature Sterilization, at 440-392-7455.
- Quantity in commerce
- 3,882 units were distributed
- Distribution
- Worldwide distribution: USA (nationwide) and countries of: Bahrain, Hong Kong, Japan and South Korea.
- Product codes and lots
- MED. Serial Numbers: 400000 through 405499 are subject to correction.