Z-0868-2012

Multi-Analyte Urine DAU Calibrator (15 mL HDPE bottle inside a cardboard box) For in vitro diagnostic use...

Lin-Zhi International Inc / initiated 2011-08-23 / Terminated / root cause: manufacturing process

Lin-Zhi International Inc began a recall of Multi-Analyte Urine DAU Calibrator (15 mL HDPE bottle inside a cardboard box) For in vitro diagnostic use only. Lin-Zhi International, Inc.. 670 Almanor Ave... on . The reason FDA records is "Multi Analyte Urine DAU Cutoff Calibrator Gives a decreased absorbance reading due to a loss of the BZO (oxazepam) analyte from the solution and may cause false positive readings". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0868-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Lin-Zhi International Inc
Product
Multi-Analyte Urine DAU Calibrator (15 mL HDPE bottle inside a cardboard box) For in vitro diagnostic use only. Lin-Zhi International, Inc.. 670 Almanor Ave. Sunnyvale, CA 94085 intended use: Drug abuse detection.
Product code
DKB Calibrators, Drug Mixture
Reason for recall
Multi Analyte Urine DAU Cutoff Calibrator Gives a decreased absorbance reading due to a loss of the BZO (oxazepam) analyte from the solution and may cause false positive readings.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm, Lin-Zhi International, Inc., sent a notice by email dated August 23, 2011 to all customers. The notice described the product, problem and actions to be taken. Lin-Zhi will contact the customers to determine whether they want replacement BZO calibrators and controls immediately or if they are willing to wait a minimum of four weeks for a new lot of the Multi-Analyte Urine Cutoff Calibrators. The customers were instructed to contact Lin-Zhi with their replacement request. The notice also stated that the Lin-Zhi has plans to create more stringent QC standards and further R&D regarding stabilization of this analyte in solution in its calibrators of controls containing BZO. If you have any questions or concerns regarding how this situation is being handled or the outcome, contact Lin-Zhi at (408)732-3856; email: [email protected] or webpage: www.lin-zhi.com.
Quantity in commerce
45 items
Distribution
Worldwide distribution: USA (nationwide) including states of: AL, FL, GA, IN, NC, NH, and TX; and country of: Germany.
Product codes and lots
Catalog number 0803; Lot 1101136K3, exp 1/28/13