Z-1067-2012

GE SIGNA EXCITE MR SURGICAL OPTION (MR450w Surgical Suite). The MR Surgical Option when integrated with the...

GE Healthcare / initiated 2011-12-15 / Terminated / root cause: manufacturing process

GE Healthcare, LLC began a recall of GE SIGNA EXCITE MR SURGICAL OPTION (MR450w Surgical Suite). The MR Surgical Option when integrated with the GE Signa 3.0T or 1.5T magnetic resonance scanner... on . The reason FDA records is "GE Healthcare has recently become aware of a potential safety issue due to the failure of the transfer board release lever associated with the OR Compatible patient transfer table of your Optima MR450w MR Surgical Suite". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1067-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
GE SIGNA EXCITE MR SURGICAL OPTION (MR450w Surgical Suite). The MR Surgical Option when integrated with the GE Signa 3.0T or 1.5T magnetic resonance scanner (previously cleared K041476) can be used to produce head and whole body magnetic resonance images that are high resolution, high signal-to-noise ratio, with short scan times. The Signa Excite MR system with Surgical Option is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. The images produced by the Signa" Excite system with Surgical Option reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis. The GE Signa" Excite system with Surgical Option can provide MR images at various stages of a surgical procedure, and provides the capability to transfer a patient to the MR system table directly from a surgical table, or patient transport system. Images can be obtained using the MR scanner body coil, or local surface coils, and the surface coils used with the integrated system may/can accommodate sterile draping for surgical procedures.
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
GE Healthcare has recently become aware of a potential safety issue due to the failure of the transfer board release lever associated with the OR Compatible patient transfer table of your Optima MR450w MR Surgical Suite.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
GE Healthcare, sent a Urgent Medical Device Correction letter dated December 15, 2011, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The customer was instructed discontinue use of OR Compatible table until issue is corrected. GE Healthcare will correct all affected systems by providing a replacement table cradle incorporating design changes to the cradle release lever mechanism at no cost to you. A GE Healthcare service representative will contact you to arrange for this correction. Contact your local service representative if you have any questions concerning this notification. For further questions please call (262) 513-4122.
Quantity in commerce
2
Distribution
Worldwide Distribution -- USA including Wisconsin and the country of Austria.
Product codes and lots
12051M69 12228M63