ThermoFisher Scientific / Microgenics brand MAS CardioImmune XL, Liquid Assayed Cardiac Marker Control, in...
Microgenics Corporation began a recall of ThermoFisher Scientific / Microgenics brand MAS CardioImmune XL, Liquid Assayed Cardiac Marker Control, in vitro diagnostic. Catalog Numbers & Description... on . The reason FDA records is "Low measurements of Troponin I in the MAS CardioImmune XL Control which were outside the published control ranges". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0637-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Microgenics Corporation
- Product
- ThermoFisher Scientific / Microgenics brand MAS CardioImmune XL, Liquid Assayed Cardiac Marker Control, in vitro diagnostic. Catalog Numbers & Description: CAI-XLL, Level L; CAI-XL1, Level 1; CAI-XL2, Level 2; CAI-XL3, Level 3; CAI-XL4, Multi-Pack; CAI-XLS, Sample Pack;
- Product code
- JJY Multi-Analyte Controls, All Kinds (Assayed)
- Reason for recall
- Low measurements of Troponin I in the MAS CardioImmune XL Control which were outside the published control ranges.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On 12/14/2011, ThermoFisher Scientific issued letters via express mail to all its direct consignees, informing them of the affected products and providing instructions on the recall.
- Quantity in commerce
- 25,113 Kits
- Distribution
- Nationwide and Worldwide, including Germany, France, Italy and Spain.
- Product codes and lots
- Catalog number, Lot Numbers & Expiration Dates: CAI-XLL, CXL1301L, 31 JAN 2013; CAI-XL1, CXL13011, 31 JAN 2013; CAI-XL2, CXL13012, 31 JAN 2013; CAI-XL3, CXL13013, 31 JAN 2013; CAI-XL4, CXL13014, 31 JAN 2013; CAI-XLS, CXL1301S, 31 JAN 2013