Siemens syngo.plaza VA20A Server Farm setup integrated to a RIS, model number 10592457
Siemens Medical Solutions USA, Inc began a recall of Siemens syngo.plaza VA20A Server Farm setup integrated to a RIS, model number 10592457 on . The reason FDA records is "The firm has become aware of the potential for an unintended merge of multiple patients when using syngo.plaza version VA20 with Server Farm setup integrated to a RIS. Sporadically it can happen that certain Patient IDs...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0827-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens syngo.plaza VA20A Server Farm setup integrated to a RIS, model number 10592457.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- The firm has become aware of the potential for an unintended merge of multiple patients when using syngo.plaza version VA20 with Server Farm setup integrated to a RIS. Sporadically it can happen that certain Patient IDs in the database are deleted and sent to NULL. This can cause incorrect merging with patients that already had a NULL Patient ID.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm initiated their product recall in May 2011 by issuing a "Customer Safety Advisory Notice" informing all customers of the malfunction and providing instructions on how to avoid the error. The issue will be solved with the next version release, planned for June 2011.
- Quantity in commerce
- 3
- Distribution
- Nationwide Distribution - Including the states of Florida, Massachusetts and Wisconsin.
- Product codes and lots
- Model number 10592457, serial numbers 100174, 100177, and 100181.