Z-0829-2012
Olympus Endoscopic Electrosurgical Unit working elements, model numbers WA22366A, and WA22367A. Working...
OLYMPUS Corporation of the Americas / initiated 2011-12-27 / Terminated / root cause: design
Olympus America Inc. began a recall of Olympus Endoscopic Electrosurgical Unit working elements, model numbers WA22366A, and WA22367A. Working element for cutting, coagulation, and vaporization in... on . The reason FDA records is "Less than optimal thickness in the plastic housing holding the electrode in the Olympus Working Elements". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0829-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- OLYMPUS Corporation of the Americas
- Product
- Olympus Endoscopic Electrosurgical Unit working elements, model numbers WA22366A, and WA22367A. Working element for cutting, coagulation, and vaporization in conductive irrigation fluid or conductive distension medium in the Saline mode.
- Product code
- KNS Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Reason for recall
- Less than optimal thickness in the plastic housing holding the electrode in the Olympus Working Elements
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Olympus America Inc. sent an "URGENT: MEDICAL DEVICE WITHDRAWAL" letter dated December 27, 2011 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to inspect their inventory for the affected product and discontinue use of any affected product discovered. Customers should contact the Gyrus ACMI Customer Service Department at 1-888-524-7266 for credit or replacement of the product. Additionally, an Olympus Working Elements Withdrawal form was enclosed for customers to complete and fax to 484-896-7128. Contact the firm at 484-896-5688 for questions regarding this notice.
- Distribution
- Nationwide Distribution-including Washington, D.C. and the states of AL, AK, AZ, AR, CA, CO, CT, FL, GA, IL, IN, IA, KS, KY, LA, ME, MD, MI, MN, MO, MT, NV, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, TN, TX, UT, VA, WA, and WI.
- Product codes and lots
- Lot numbers 118W, 119W, and 11XW for new inventory
FIRST ARTICLE / WEEKLY