Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1450-2012

Sarns Modular Perfusion System 8000 Catalog number 16414 Cardioplegia Monitor The 8K Arterial Monitor is...

Terumo Cardiovascular Systems Corporation / initiated 2012-03-21 / Terminated / root cause: design

Terumo Cardiovascular Systems Corporation began a recall of Sarns Modular Perfusion System 8000 Catalog number 16414 Cardioplegia Monitor The 8K Arterial Monitor is intended for use with the 8000 Modular Perfusion... on . The reason FDA records is "Terumo received multiple reports of malfunctions of the Cardioplegia and arterial monitors for the Sarns Modular Perfusion System 8000. Upon review of complaint investigation, the malfunctions are caused by or could...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1450-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Sarns Modular Perfusion System 8000 Catalog number 16414 Cardioplegia Monitor The 8K Arterial Monitor is intended for use with the 8000 Modular Perfusion System to display one average pressure and three average temperatures and also provide two digital timers. The pressure channel can have alert and alarm levels set by the user to indicate overpressure conditions. The alert and alarm level settings are retained by the system and can be recalled after the next power up. Alert and alarm conditions are indicated by flashing LED bar with single or dual audio tones; an alarm will also stop the arterial and cardioplegia pumps.
Product code
DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Reason for recall
Terumo received multiple reports of malfunctions of the Cardioplegia and arterial monitors for the Sarns Modular Perfusion System 8000. Upon review of complaint investigation, the malfunctions are caused by or could have been caused by soldering anomalies on the circuit board of the monitor. The compromised solder joints could cause intermittent or loss of display, or could cause monitor reset, po
Root cause tag
design
FDA root cause
Device Design
Action
TERUMO sent an URGENT MEDICAL DEVICE RECALL- SAFETY ADVISORY to all consignees on March 21, 2012 via Federal Express. Terumo issued the safety advisory to remind users how to intervene if they experienced intermittent or complete loss of pressure display on the arterial monitor or cartiplegia monitor, or a related pump stop. Customers were asked to review the Medical Device Correction notice, assure that all users are aware of the notice, confirm receipt of the communication by faxing the attached Customer Response Form to 1-800-292-6551. For questions regarding this recall call 1-800-521-2818.
Quantity in commerce
3,022 units
Distribution
Worldwide Distribution - USA including AK, AL , AR, AZ , CA, CO , CT, DC, DE , FL, GA, HI, IA , ID , IL, IN , KS, KY , LA, MA , MD, MI, MN , MO, MS , MT, NC , ND, NE, NH, NJ, NM , NY, OH , OR, PA , RI, SC , TN, TX , VA, WA , WI, WV, WY, and Puerto Rico and the countries of PERU, Philippines, REPUBLICA, SANTIAGO, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Tujunga, United Arab Emirates, Uruguay, Venezuela, Vietnam, Wooddale, and Yemen.
Product codes and lots
Catalog Number 16414; Lot/serial numbers: PT001*, 1001-1064, 1070-1146, 1148-1220,1222-1227,2000-2039,2045-2059,2061- 2064,2100-2157,2159-2203,2205-2335,2337-2399, 2401-2408,3000-3014,3016-3070,3072-3110,3112-3385, 3387-3558, and 3560-3840.