Sarns Modular Perfusion System 8000 Catalog number 16413 Arterial Monitor The 8K Arterial Monitor is intended...
Terumo Cardiovascular Systems Corporation began a recall of Sarns Modular Perfusion System 8000 Catalog number 16413 Arterial Monitor The 8K Arterial Monitor is intended for use with the 8000 Modular Perfusion System to... on . The reason FDA records is "Terumo received multiple reports of malfunctions of the Cardioplegia and arterial monitors for the Sarns Modular Perfusion System 8000. Upon review of complaint investigation, the malfunctions are caused by or could...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1451-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- Sarns Modular Perfusion System 8000 Catalog number 16413 Arterial Monitor The 8K Arterial Monitor is intended for use with the 8000 Modular Perfusion System to display one average pressure and three average temperatures and also provide two digital timers. The pressure channel can have alert and alarm levels set by the user to indicate overpressure conditions. The alert and alarm level settings are retained by the system and can be recalled after the next power up. Alert and alarm conditions are indicated by flashing LED bar with single or dual audio tones; an alarm will also stop the arterial and cardioplegia pumps.
- Product code
- DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
- Reason for recall
- Terumo received multiple reports of malfunctions of the Cardioplegia and arterial monitors for the Sarns Modular Perfusion System 8000. Upon review of complaint investigation, the malfunctions are caused by or could have been caused by soldering anomalies on the circuit board of the monitor. The compromised solder joints could cause intermittent or loss of display, or could cause monitor reset, po
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- TERUMO sent an URGENT MEDICAL DEVICE RECALL- SAFETY ADVISORY to all consignees on March 21, 2012 via Federal Express. Terumo issued the safety advisory to remind users how to intervene if they experienced intermittent or complete loss of pressure display on the arterial monitor or cartiplegia monitor, or a related pump stop. Customers were asked to review the Medical Device Correction notice, assure that all users are aware of the notice, confirm receipt of the communication by faxing the attached Customer Response Form to 1-800-292-6551. For questions regarding this recall call 1-800-521-2818.
- Quantity in commerce
- 3,022
- Distribution
- Worldwide Distribution - USA including AK, AL , AR, AZ , CA, CO , CT, DC, DE , FL, GA, HI, IA , ID , IL, IN , KS, KY , LA, MA , MD, MI, MN , MO, MS , MT, NC , ND, NE, NH, NJ, NM , NY, OH , OR, PA , RI, SC , TN, TX , VA, WA , WI, WV, WY, and Puerto Rico and the countries of PERU, Philippines, REPUBLICA, SANTIAGO, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Tujunga, United Arab Emirates, Uruguay, Venezuela, Vietnam, Wooddale, and Yemen.
- Product codes and lots
- Catalog Number 16413; Lot numbers/serial numbers: PT003*, 1001-1079, 1085-1270, 1276,2000-2013,2015-2028, 2030-2109, 2111-2295, 2297-2396, 2398-2433, 2435-2439, 3000-3020, 3022-3052, 3054-3065, 3067-3104, 3106-3197, 3199-3211, 3213-3593, 3595-3634, 3636-3827, and 3829-3908.