Z-0818-2012
EB- Series Video Bronchoscopes, EB-1570K & EB-1970K. Intended to provide optical visualization of (via a...
PENTAX of America, Inc. / initiated 2011-09-30 / Terminated / root cause: manufacturing process
Pentax Medical Company began a recall of EB- Series Video Bronchoscopes, EB-1570K & EB-1970K. Intended to provide optical visualization of (via a video monitor), and therapeutic access to, the... on . The reason FDA records is "The wrong silicone sealant was used and excess sealant was neglected to be removed during repair of the bronchoscope". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0818-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- PENTAX of America, Inc.
- Product
- EB- Series Video Bronchoscopes, EB-1570K & EB-1970K. Intended to provide optical visualization of (via a video monitor), and therapeutic access to, the Pulmonary Tract.
- Product code
- EOQ Bronchoscope (Flexible Or Rigid)
- Reason for recall
- The wrong silicone sealant was used and excess sealant was neglected to be removed during repair of the bronchoscope.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Pentax sent an "Urgent: Device Recall - Stop Use Immediately" letter/return response form to their customers on 9/30/2011. The letter identified the affected product and asked customers to immediately discontinue the use of their endoscope(s). Customers were to contact PENTAX at 1-800-736-8293 to schedule the return of the affected product(s). The reason for the recall was also explained, and the letter stated that there will be no charge for shipping, evaluation, or repair of the products. Customers can request loaner instruments, if necessary.
- Quantity in commerce
- 18 units
- Distribution
- Nationwide Distribution.
- Product codes and lots
- 1) EB-1570K G110999 A110254 I A110067 G110338 G110522 G111115 A110125 A110019 A110277 G110538 G111192 G111270 G111216 G111254 G110973 A110022 G111250 2) EB-1970K A110151