Z-0860-2012

Siemens AXIOM Aristos FX Plus solid state x-ray imager. Image-intensified fluoroscopic x-ray system

Siemens Medical Solutions USA, Inc. / initiated 2011-12-12 / Terminated / root cause: manufacturing process

Siemens Medical Solutions USA, Inc began a recall of Siemens AXIOM Aristos FX Plus solid state x-ray imager. Image-intensified fluoroscopic x-ray system on . The reason FDA records is "Fixation bolts for the mounting of the X-ray tube holder became mixed with an incorrect bolt type". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0860-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens AXIOM Aristos FX Plus solid state x-ray imager. Image-intensified fluoroscopic x-ray system.
Product code
MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)
Reason for recall
Fixation bolts for the mounting of the X-ray tube holder became mixed with an incorrect bolt type.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Siemens Healthcare sent a "SAFETY ADVISORY NOTICE" dated December 12, 2011 to the affected customer. The letter identifies the product, problem, and actions to be taken by the customer. The firm prepared a modification, which will be provided to the affected customer under Update Instruction XP055/11/S. The firm will resolve this potential issue by preventively exchanging all bolts for the fixation of the X-ray tube holder of the concerned AXIOM Aristos FX Plus unit.
Quantity in commerce
1
Distribution
Product distributed in one hospital in California.
Product codes and lots
Serial number 1273