Z-0743-2012

Affected device is Model 98400-SL6-AHA, CardioExpress Electrocardiograph. The ECG cable part number is...

GE Healthcare / initiated 2010-09-10 / Terminated / root cause: manufacturing process

Del Mar Reynolds Medical, Ltd. began a recall of Affected device is Model 98400-SL6-AHA, CardioExpress Electrocardiograph. The ECG cable part number is 01.57.471017. The electrocardiograph is only intended to... on . The reason FDA records is "CardioExpress SL6 ECG cable provided with the electrocardiograph is mislabeled. The left arm (LA) is mislabeled as right leg (RL). The left leg (LL) is mislabeled as left arm (LA). The right leg (RL) is mislabeled as...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0743-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
Affected device is Model 98400-SL6-AHA, CardioExpress Electrocardiograph. The ECG cable part number is 01.57.471017. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The intended use of electrocardiograph is to acquire ECG signals from adult and pediatric patients through body surface with ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is only intended to be used on adult patients and is offered to clinicians on an advisory basis only.
Product code
DPS Electrocardiograph
Reason for recall
CardioExpress SL6 ECG cable provided with the electrocardiograph is mislabeled. The left arm (LA) is mislabeled as right leg (RL). The left leg (LL) is mislabeled as left arm (LA). The right leg (RL) is mislabeled as left leg (LL).
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Spacelabs Healthcare sent an "URGENT-MEDICAL DEVICE CORRECTION" letter dated September 24, 2010 to one affected customer in Massachusetts. The letter described the product, problem, and actions to be taken by the customers. Customers were instructed to review previous diagnoses made from this electrocardiograph to mitigate any inappropriate recordings or treatment made, and dispose of the mislabeled ECG cable to prevent its use. Spacelabs replaced the affected ECG cable with the new ECG cable. If you have any questions about this corrective action program, please contact Spacelabs Healthcare at 1-800-522-7025 and select 2 for Technical Support.
Quantity in commerce
15 units
Distribution
Nationwide distribution.
Product codes and lots
Serial numbers:  98400-SL6-AHA1030508, 98400-SL6-AHA1030509, 98400-SL6-AHA1030510, 98400-SL6-AHA1030511, 98400-SL6-AHA1030512, 98400-SL6-AHA1030513, 98400-SL6-AHA1030514, 98400-SL6-AHA1030515, 98400-SL6-AHA1030516, 98400-SL6-AHA1030517, 98400-SL6-AHA1030518, 98400-SL6-AHA1030519, 98400-SL6-AHA1030520, 98400-SL6-AHA1030521, 98400-SL6-AHA1030522.