Z-0721-2012
Angiodynamics DuraMax 15.5F x 28 cm Chronic Hemodialysis Catheter Set (with Cuff 23cm from Tip) --- Catalog...
Angiodynamics Worldwide Headquarters / initiated 2011-12-30 / Terminated / root cause: manufacturing process
Angiodynamics Worldwide Headquarters began a recall of Angiodynamics DuraMax 15.5F x 28 cm Chronic Hemodialysis Catheter Set (with Cuff 23cm from Tip) --- Catalog No. [REF] 10302804 --- MANUFACTURED IN USA... on . The reason FDA records is "The securement cuff on the hemodialysis catheter has the potential to detach from the shaft". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0721-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Angiodynamics Worldwide Headquarters
- Product
- Angiodynamics DuraMax 15.5F x 28 cm Chronic Hemodialysis Catheter Set (with Cuff 23cm from Tip) --- Catalog No. [REF] 10302804 --- MANUFACTURED IN USA AngioDynamics, Inc., 603 Queensbury Avenue, Queensbury, NY USA 12804 Product Usage: indicated for use in attaining long-term vascular access for Hemodialysis and Apheresis.
- Product code
- MSD Catheter, Hemodialysis, Implanted
- Reason for recall
- The securement cuff on the hemodialysis catheter has the potential to detach from the shaft.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- AngioDynamics sent an Urgent - Medical Device Recall letter dated December 30, 2011, via certified mail to all affected customers. The letter identified the affected product, the problem, and the actions to be taken. Customers were instructed to examine inventory, segregate the recalled lots and return remaining inventory to the firm. The letter states that replacement product will be shipped upon receipt and confirmation of the returned product. Customers were advised to complete the Recall Reply Form included and fax to DuraMax Hemodialysis Catheter Recall Coordinator at 518-798-1360. For questions contact your local representative or call the Customer Service Manger at ANGIODYNAMICS Customer Service at 1-800-772-6446.
- Quantity in commerce
- 1336 units
- Distribution
- Worldwide Distribution - USA (Nationwide)
- Product codes and lots
- FG Lot Numbers: 541558, 542262, 542266, 543822, 543823, 556670, 556672, C40503