Visum LED Surgical Lights, Ceiling Mounted Surgical Light. The Visum LED surgical light is intended to...
Stryker Communications Corp began a recall of Visum LED Surgical Lights, Ceiling Mounted Surgical Light. The Visum LED surgical light is intended to illuminate the operative site during surgical procedures... on . The reason FDA records is "Stryker has identified a potential structural issue with the Visum LED surgical light(s). During a complaint investigation, Stryker became aware that there was a possibility that incorrect assembly could result in the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0913-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- STRYKER Communications, Inc.
- Product
- Visum LED Surgical Lights, Ceiling Mounted Surgical Light. The Visum LED surgical light is intended to illuminate the operative site during surgical procedures with high intensity light.
- Product code
- FSY Light, Surgical, Ceiling Mounted
- Reason for recall
- Stryker has identified a potential structural issue with the Visum LED surgical light(s). During a complaint investigation, Stryker became aware that there was a possibility that incorrect assembly could result in the light head beginning to separate from the suspension arm. In the event the light separates and falls from the arm, it could lead to serious injury.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The firm sent "Urgent: Medical Device Correction" letters dated 11/07/2011 to their domestic and international customers. The letter informed customers of the remote risk of the light detaching from the arm and recommended continued use of the product. A representative will contact them about scheduling an appointment to inspect the lights. Contact Technical Services at 1-866-841-5663 for questions.
- Quantity in commerce
- 1143 Devices
- Distribution
- Worldwide Distribution - USA, and the countries of Australia, Canada, Hong Kong, Brazil, Sweden, India, Japan, Korea, Ecuador, Venezuela, Mexico, Denmark, Great Britain, and New Zealand.
- Product codes and lots
- Catalog Number: 0682001250, 0682001251, 0682001252, and 0682001253. Product Numbers: P10235 (Visum LED Surgical Light, Camera Ready), P10234 (Visum LED Surgical Light), P12875 (Visum LED Surgical Light Camera Ready - Low Ceiling), P12874 (Visum LED Surgical Light - Low Ceiling). Expiration date: There is no expiration date with the scope of the action. Serial Numbers: Product manufactured between October 2010 to July 2011 and in use by customer.