Z-0725-2012

Angiodynamics DuraMax VP Vascpak Catheter Kit 15.5F x 22cm Chronic Hemodialysis VascPak (with Cuff 17cm from...

Angiodynamics Worldwide Headquarters / initiated 2011-12-30 / Terminated / root cause: manufacturing process

Angiodynamics Worldwide Headquarters began a recall of Angiodynamics DuraMax VP Vascpak Catheter Kit 15.5F x 22cm Chronic Hemodialysis VascPak (with Cuff 17cm from Tip) --- Catalog No. [REF] 10302817 ---... on . The reason FDA records is "The securement cuff on the hemodialysis catheter has the potential to detach from the shaft". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0725-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Angiodynamics Worldwide Headquarters
Product
Angiodynamics DuraMax VP Vascpak Catheter Kit 15.5F x 22cm Chronic Hemodialysis VascPak (with Cuff 17cm from Tip) --- Catalog No. [REF] 10302817 --- MANUFACTURED IN USA AngioDynamics, Inc., 603 Queensbury Avenue, Queensbury, NY USA 12804 Product Usage: indicated for use in attaining long-term vascular access for Hemodialysis and Apheresis.
Product code
MSD Catheter, Hemodialysis, Implanted
Reason for recall
The securement cuff on the hemodialysis catheter has the potential to detach from the shaft.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
AngioDynamics sent an Urgent - Medical Device Recall letter dated December 30, 2011, via certified mail to all affected customers. The letter identified the affected product, the problem, and the actions to be taken. Customers were instructed to examine inventory, segregate the recalled lots and return remaining inventory to the firm. The letter states that replacement product will be shipped upon receipt and confirmation of the returned product. Customers were advised to complete the Recall Reply Form included and fax to DuraMax Hemodialysis Catheter Recall Coordinator at 518-798-1360. For questions contact your local representative or call the Customer Service Manger at ANGIODYNAMICS Customer Service at 1-800-772-6446.
Distribution
Worldwide Distribution - USA (Nationwide)
Product codes and lots
FG Lot Numbers: C37756