Z-1094-2012
ORIGIO MIDATLANTIC DEVICES CRYOPETTE (r) Rx only Single use only CRY-PETTE-CL LOT 123456 2013-10-10 Exp. date...
CooperSurgical, Inc. / initiated 2011-11-28 / Terminated / root cause: design
ORIGIO HUMAGEN PIPETS INC. began a recall of ORIGIO MIDATLANTIC DEVICES CRYOPETTE (r) Rx only Single use only CRY-PETTE-CL LOT 123456 Exp. date *** MADE IN THE USA STERILE R ***Origio, Inc... on . The reason FDA records is "Cryopettes were rupturing during the warming process following vitrification". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1094-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- CooperSurgical, Inc.
- Product
- ORIGIO MIDATLANTIC DEVICES CRYOPETTE (r) Rx only Single use only CRY-PETTE-CL LOT 123456 2013-10-10 Exp. date *** MADE IN THE USA STERILE R ***Origio, Inc. 3400 Hunter's Way, Charlottesville, VA 22911 www.origio.com The Cryopette (r) is a cryopreservative storage device that is intended for use in vitrification procedures to contain and maintain human blastocyst stage embryos.
- Product code
- MQK Labware, Assisted Reproduction
- Reason for recall
- Cryopettes were rupturing during the warming process following vitrification.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- ORIGIO sent a Field Safety Corrective Action letter dated November 21, 2011, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. ORIGIO feels it is important to eliminate this risk completely. As soon as we are confident that this danger of rupture has been eliminated, we will make Cryopettes available again. Please use the attached UPS call tag to return any unused Cryopettes to ORIGIO Inc. Upon receipt, we will issue a credit to your account. We sincerely apologize for the problems this issue has caused in your laboratories. For further questions please call (434) 979-4000.
- Quantity in commerce
- 73850 Cryopettes
- Distribution
- Worldwide Distribution - U.S. included states of AL, AR, DE, PR, NY, MA, NH, VT, IL, CT, NJ, PA, VA, WV, NV, NC, SC, GA, FL, OH, MD, MS, IN, MI, PR, WI, MN, SD, MT, MO, TX, CO, UT, AZ, CA, HI, OR, AK, WA and in countries of Turkey, Saudi Arabia, Kuwait, Jordon, So. Korea, Taiwan, Brazil, So. Africa, Oman, Kazakhstan, Honduras, Slovenia, Serbia, Poland, Czech Republic, Romania, Croatia, Ukraine, Netherlands, Bulgaria, Hungary, Russia, Lebanon, Singapore, Malaysia, Vietnam, China, Thailand, Nigeria, Japan, Spain, Greece, Canada, Denmark, France, United Kingdom, Italy, Germany, and Australia,
- Product codes and lots
- Lot numbers/Expiration: 27109A exp. 9/2011; 29909A, exp. 10/2011; 06210A, exp. 3/2012; 06210B, exp. 3/3/2012; 16210A, exp. 6/2012; 20810A, exp. 7/2012; 22410A, exp. 8/2012; 25610A, exp. 9/2012; 32010A, exp. 11/2012; 32010B, exp. 11/2012; 35510A, exp. 12/2012; 03210A, exp. 2/2012; 05411A, exp. 2/2013; 09811C, exp. 4/2013; 09811A, exp. 4/2013; 09811B, exp. 4/2013; 14611A, exp. 5/2013; 14611B, exp. 2013; 020111, exp. 8/1/13; 020111, exp. 8/8/13; 020111, exp. 8/15/13; 020111, exp. 9/5/13; 020111, exp. 9/12/13; 020111, exp. 9/26/13