Z-1011-2012

Zoll pedi.padz Reduced Energy Multi-Function Electrodes (MFE) Part number: 8900-0401 Used in connection with...

ZOLL Medical Corporation / initiated 2011-12-30 / Terminated / root cause: manufacturing process

Bio-Detek, Inc. began a recall of Zoll pedi.padz Reduced Energy Multi-Function Electrodes (MFE) Part number: 8900-0401 Used in connection with Zoll M Series and E series Defibrillators... on . The reason FDA records is "Lack of an adequate silicone seal around the periphery of the attenuator enclosure may allow fluid to enter". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1011-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ZOLL Medical Corporation
Product
Zoll pedi.padz Reduced Energy Multi-Function Electrodes (MFE) Part number: 8900-0401 Used in connection with Zoll M Series and E series Defibrillators specifically for pediatric patients.
Reason for recall
Lack of an adequate silicone seal around the periphery of the attenuator enclosure may allow fluid to enter.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Zoll Medical issued Urgent Device Correction notices dated 12/30/11 via UPS/Registered Mail. The letter described the issue and identified the affected product. ZOLL recommended that customers alert their staff of the issue, that the electrodes should be kept away from a fluid environment, to use caution in handling the electrodes if they are subjected to a fluid environment and maintain safe defibrillation practice, and seek a replacement if available. Also customers should contact Zoll 1-800-348-9011 for product replacement. The firm plans to replace the affected lots free of charge.
Quantity in commerce
3,080 units
Distribution
Worldwide Distribution -- US and Canada.
Product codes and lots
Lot Numbers 0510 thru 4511