Z-0738-2012
Linet Remote Handset for model Multicare, Eleganza 3 and Eleganza Smart, hospital bed
Linet Americas / initiated 2011-12-08 / Terminated / root cause: manufacturing process
Linet Americas Inc began a recall of Linet Remote Handset for model Multicare, Eleganza 3 and Eleganza Smart, hospital bed. Product Usage: Positional hospital bed for acute care. The bed is... on . The reason FDA records is "The premature failure of the Remote Handset buttons may result in unintended movement of the bed". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0738-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Linet Americas
- Product
- Linet Remote Handset for model Multicare, Eleganza 3 and Eleganza Smart, hospital bed. Product Usage: Positional hospital bed for acute care. The bed is designed for the care and treatment of inpatients at hospitals and is designed for standard and intensive care.
- Product code
- FNL Bed, Ac-Powered Adjustable Hospital
- Reason for recall
- The premature failure of the Remote Handset buttons may result in unintended movement of the bed.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Linet sent a notification letter dated December 9, 2011 to all affected customers. The letter identified the product, problem, and actions to be taken. The letter states that firm Linet has decided to do a voluntary field correction to replace the recalled product. The replacement will be completed by the territory Representative (Linet Employee) who will replace the affected product for each customer personally. The replacement takes approximately 30 seconds and does not require the patient to be moved. For questions or concerns call 704-248-5650.
- Quantity in commerce
- 29,252 units
- Distribution
- Worldwide Distribution - USA (nationwide) including the states of: AL, FL, KY, LA, MA, NC, NY, OH and the country of: Canada
- Product codes and lots
- Model numbers: 1GZ, 1GV, IS, 1GM, 1AG, 1MC and 1L.