Z-0817-2012

Hospira Blood Set, with 200 Micron Filter, 80 Inch, Non-DEHP; an Rx sterile medical device used to administer...

Abbott Medical / initiated 2012-01-12 / Terminated / root cause: design

Hospira Inc. began a recall of Hospira Blood Set, with 200 Micron Filter, 80 Inch, Non-DEHP; an Rx sterile medical device used to administer blood products; 48 individually wrapped sets per... on . The reason FDA records is "Silicone applied to the piercing pin of the blood set is reducing retention force of the piercing pin while in use. This could cause the piercing pin to slip out of the administration container". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0817-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Hospira Blood Set, with 200 Micron Filter, 80 Inch, Non-DEHP; an Rx sterile medical device used to administer blood products; 48 individually wrapped sets per case; Made in Costa Rica, Hospira, Inc., Lake Forest, IL 60045 USA; list 14203-28. Intended use: for the administration of blood products.
Product code
FPA Set, Administration, Intravascular
Reason for recall
Silicone applied to the piercing pin of the blood set is reducing retention force of the piercing pin while in use. This could cause the piercing pin to slip out of the administration container.
Root cause tag
design
FDA root cause
Component design/selection
Action
The firm, Hospira, sent an "URGENT DEVICE RECALL" letter dated January 12, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to check their inventory and immediately quarantine any affected product. Additionally, a Reply Form was attached for customers to complete and return via fax to 1-866-912-8410 or e-mail to: [email protected]. Please contact Hospira Customer Care at 1-877-946-7747 for questions regarding this recall.
Quantity in commerce
960 units
Distribution
Distributed in the states of North Carolina and Washington.
Product codes and lots
list 14203-28, lot 96-146-5H