Microsoft Amalga - previously known as Microsoft Amalga Unified Intelligence System (UIS) 2009
Microsoft Corporation Health Solutions Group began a recall of Microsoft Amalga - previously known as Microsoft Amalga Unified Intelligence System (UIS) 2009 Product Usage: Microsoft Amalga UIS 2009 enables healthcare... on . The reason FDA records is "Healthcare providers reported that when sharing a single browser session on Amalga Web, patient A's data can be displayed under patient B's name". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1082-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Microsoft Corporation Health Solutions Group
- Product
- Microsoft Amalga - previously known as Microsoft Amalga Unified Intelligence System (UIS) 2009 Product Usage: Microsoft Amalga UIS 2009 enables healthcare providers to create a comprehensive continually updated data ecosystem to aggregate, store, display, and share disparate data either coming from many source systems or directly entered by end users for rapid self-service data exploration and analysis by front line managers, clinicians and ancillary staff. Furthermore, Amalga UIS allows healthcare providers to rapidly create new applications to respond to evolving business and clinical needs.
- Product code
- JQP Calculator/Data Processing Module, For Clinical Use
- Reason for recall
- Healthcare providers reported that when sharing a single browser session on Amalga Web, patient A's data can be displayed under patient B's name.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Microsoft Corporation sent an "Urgent Field Safety Notice" letter dated January 5, 2012 by email to all affected customers. The letter described the product, problem, and actions to be taken by the customers. If you have any questions about this corrective action program, please contact Microsoft Corporation at 425-882-8080.
- Quantity in commerce
- 9 units
- Distribution
- Worldwide Distribution - USA (nationwide) including the states of California, China, Kentucky, North Carolina, Wisconsin and country of Switzerland.
- Product codes and lots
- Software Version: R2 SP2, RTM, R2 SP3