Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0914-2012

enGen(TM) Laboratory Automation Systems configured with Hettich Rotanta 46 RSC Robotic Centrifuges, Catalog #...

Ortho-Clinical Diagnostics, Inc. / initiated 2011-04-28 / Terminated / root cause: design

Ortho-Clinical Diagnostics began a recall of enGen(TM) Laboratory Automation Systems configured with Hettich Rotanta 46 RSC Robotic Centrifuges, Catalog # enGen -- Common/Usual Name: enGen(TM) Track... on . The reason FDA records is "Ortho Clinical Diagnostics received complaints of centrifuge buckets disconnecting from their supports during centrifugation when using the Hettich Rotanta 46 RSC Robotic Centrifuges, Models 4815, 4816 and 4817, that...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0914-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
enGen(TM) Laboratory Automation Systems configured with Hettich Rotanta 46 RSC Robotic Centrifuges, Catalog # enGen -- Common/Usual Name: enGen(TM) Track System, Hettich Centrifuge -- Ortho-Clinical Diagnostics, Inc. Rochester, NY --- For in vitro diagnostic use only. The VITROS 5,1 FS Chemistry System with enGen(TM) Laboratory Automation System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods).
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Ortho Clinical Diagnostics received complaints of centrifuge buckets disconnecting from their supports during centrifugation when using the Hettich Rotanta 46 RSC Robotic Centrifuges, Models 4815, 4816 and 4817, that may be configured with enGen(TM) Laboratory Automation Systems.
Root cause tag
design
FDA root cause
Component design/selection
Action
Ortho-Clinical Diagnostics sent an Urgent Product Correction Notification letter dated April 28, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to post the enclosed attachment near their centrifuge as a reference. Clean and carefully inspect the suspension groove area of the centrifuge buckets for cracks as part of their weekly maintenance procedure. Careful to check for micro-cracks that may not be easily detectable. If any cracks/micro-cracks, corrosion or damage are present, they must immediately discontinue using the centrifuge and contact OCD Customer Technical Services at 1-800-421-3311 to obtain replacements. OCD recommends that they do not exceed the manufacturer's relative centrifuge force (RCF) and run time limits for the blood collection devices used by their laboratory. Complete and return the attached Confirmation of Receipt Form. For questions regarding this recall call the Customer Service Center at 1-800-421-3311. .
Quantity in commerce
78 units worldwide (18 Domestic and 68 Foreign)
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Australia, Canada, Chile, England, France, Germany, Italy, Mexico, Singapore and Spain.
Product codes and lots
All enGen(TM) Laboratory Automation Systems configured with Hettich Rotanta 46 RSC Robotic Centrifuges.