Z-0919-2012

The Artificial Bur Patch is a Temporary Fascial Prothesis. For Single Patient Use only. Use with Negative...

Novomedicus LLC / initiated 2011-06-22 / Terminated

Novomedicus Llc began a recall of The Artificial Bur Patch is a Temporary Fascial Prothesis. For Single Patient Use only. Use with Negative Pressure Wound Dressing. Contents: Large Bur Patch... on . The reason FDA records is "On 06/22/2011 Novomedicus, FL initiated a recall of the Artificial Bur Hypopack Lot 1010.1, 1102.1, 1103.1 and 5643.2. Cytotoxicity testing revealed a biocompatibility issue. The test results did not represent a health...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0919-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Novomedicus LLC
Product
The Artificial Bur Patch is a Temporary Fascial Prothesis. For Single Patient Use only. Use with Negative Pressure Wound Dressing. Contents: Large Bur Patch (Sterile: EO), 20cm x 40cm Loop Sheet, 20cm x 40cm Hook Sheet. Sterile: Contents are sterile for maximum two years unless enclosed package has been opened or damaged. Manufactured For: NOVOMEDICUS, LLC P.O. Box 2105 Nokomis, FL 34274. Used for a fascia expander prosthesis to temporarily close the abdominal cavity when regular abdominal closure is not possible; to prevent perfusion impairment of abdominal organs from increases intra-abdominal pressure; and for advanced intra-abdominal pathology requiring more than one abdominal operation to control the disease process.
Product code
FTL Mesh, Surgical, Polymeric
Reason for recall
On 06/22/2011 Novomedicus, FL initiated a recall of the Artificial Bur Hypopack Lot 1010.1, 1102.1, 1103.1 and 5643.2. Cytotoxicity testing revealed a biocompatibility issue. The test results did not represent a health risk.
FDA root cause
Process change control
Action
Novomedicus sent a Medical Device Correction letter dated June 22, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to destroy the inventory, or return the inventory to Novomedicus. If customers decided to return the remaining inventory, Novomedicus would provide TWO replacement kits for each one returned at no additional cost. Customers were asked to complete the attached Return Response form and return to NovoMedicus using the self-addressed and stamped envelope included. For questions regarding this recall call 941-375-4108.
Quantity in commerce
10 patches
Distribution
Wordwide Distribution - USA including IL, AZ, PA, CT, MO, WI, NY, TX, OR, ME, FL, VA, and MS and the countries of Mexico and Qatar.
Product codes and lots
Lot 54643.2