Z-0985-2012

MultiDiagnost Eleva IITV (MD-EII) with Software Release: PBL 6.1.1 Product Code 708036

Philips North America LLC / initiated 2012-01-09 / Terminated / root cause: manufacturing process

Philips Healthcare Inc. began a recall of MultiDiagnost Eleva IITV (MD-EII) with Software Release: PBL 6.1.1 Product Code 708036 Product Usage: The Philips MultiDiagnost Eleva system is intended to be... on . The reason FDA records is "Incorrect Skin Dose calculation. AKR (Air Kerma Rate) displayed and stored in patient records is too low". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0985-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
MultiDiagnost Eleva IITV (MD-EII) with Software Release: PBL 6.1.1 Product Code 708036 Product Usage: The Philips MultiDiagnost Eleva system is intended to be used as a multifunctional/universal system. General R/F, Fluoroscopy, Radiography and Angiography can be performed along with more specialized interventional applications. This includes the following general areas: Digestive system, Skeletal system, Urinary system, Reproductive system, Respiratory system, Circulatory system, and Various: Arthrograms, Myelograms, Facet joint injections, Discography, Sialography.
Product code
IZI System, X-Ray, Angiographic
Reason for recall
Incorrect Skin Dose calculation. AKR (Air Kerma Rate) displayed and stored in patient records is too low
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Philips Healthcare issued a mandatory Urgent Field Safety Noticedated October 3, 2011 to all affected customers. The letter identifies the affected product, problem description, hazard involved, action to be taken by customer/user and actions planned by Philips to correct the problem. The letter provide important information for the continued safe and proper use of the equipment. The letter instructs customers to review the following information with all members of their staff who need to be aware of the contents of this communication and retain a copy with the equipment Instruction for Use. For further information or support concerning this issue contact your local Philips representative at 866-767-2822.
Quantity in commerce
10 units (US)
Distribution
Worldwide Distribution - USA (nationwide) including the states of CA, IL, OH, SD, TX, WY and the countries of: Australia, Austria, Canada, Egypt, Germany, mexico, Turkey and UK
Product codes and lots
Software Release: PBL 6.1.1